Z-0429-2026Class II

Beckman Coulter UniCel DxI 800, Part Numbers: 973100, A71456; Photometric Chemistry Analyzer

Siemens Healthcare Diagnostics Inc. / initiated 2025-10-01 / Open, Classified / root cause: manufacturing process

Beckman Coulter, Inc. began a recall of Beckman Coulter UniCel DxI 800, Part Numbers: 973100, A71456; Photometric Chemistry Analyzer on . FDA classified it as Class II and gives the reason as "An internal investigation has determined that due to defective sub-components within the gantry assembly, a z-axis CAD movement motion error may occur and cause an analyzer to enter a Not Ready or Paused state, leading...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0429-2026
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
Beckman Coulter UniCel DxI 800, Part Numbers: 973100, A71456; Photometric Chemistry Analyzer
Product code
JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Reason for recall
An internal investigation has determined that due to defective sub-components within the gantry assembly, a z-axis CAD movement motion error may occur and cause an analyzer to enter a Not Ready or Paused state, leading to a delay in obtaining results.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Beckman Coulter issue an URGENT MEDICAL DEVICE RECALL notice to its consignees on 10/01/2025 via USPS and email. The notice explained the issue, potential impact, and requested the following: If Wash PnP and/or Incubator PnP Z-motion errors appear in the analyzer event log, take the following steps: 1. Attempt to reinitialize the analyzer: a. Go to the Diagnostics screen on the User Interface. b. Select Initialize System. 2. If reinitialization is unsuccessful, contact Beckman Coulter Technical Support in your region. Beckman Coulter recommends posting this letter on or near the affected analyzers. Please share this information with your laboratory staff and retain this notification as part of your laboratory Quality System documentation. If you have forwarded any of the affected products listed above to another laboratory, please provide them a copy of this letter. A field service representative will contact your laboratory to schedule a time to replace the affected parts. If you have any questions regarding this notice, please contact our Customer Support Center website: http://www.beckmancoulter.com
Quantity in commerce
657 units
Distribution
Worldwide
Product codes and lots
973100, UDI/DI 15099590265373 (Chaska), Serial numbers: 610458, 610455, 610679, 610579, 610432, 610430, 610434, 610664, 610682, 610765, 610500, 610541, 610544, 610545, 610562, 610547, 610585, 610595, 610475, 610453, 610781, 610580, 610444, 610704, 610586, 610684, 610589, 610588, 610493, 610659, 610570, 610498, 610748, 610496, 610560, 610573, 610574, 610577, 610601, 610651, 610680, 610642, 610641, 610683, 610747, 610468, 610772, 610441, 610476, 610489, 610488, 610492, 610532, 610534, 610533, 610535, 610640, 610457, 610464, 610790, 610559, 610687, 610571, 610552, 610555, 610582, 610531, 610568, 610594, 610773, 610490, 610668, 610784, 610812, 610556, 610550, 610600, 610565, 610558, 610557, 610551, 610527, 610685, 610730, 610686, 610494, 610753, 610590, 610593, 610546, 610724, 610538, 610540, 610439, 610491, 610581, 610649, 610466, 610789, 610723, 610575, 610443, 610776, 610734, 610518, 610519, 610517, 610520, 610563, 610569, 610787, 610459, 610775, 610572, 610561, 610597, 610445, 610440, 610567, 610596, 610497, 610774, 610454, 610426, 610431, 610442, 610433, 610435, 610539, 610578, 610509, 610418, 610483, 610484, 610515, 610425, 610503, 610510, 610452, 610526, 610602, 610481, 610487, 610482, 610507, 610505, 610506, 610514, 610499, 610504, 610511, 610423, 610524, 610525, 610512, 610564, 610424, 610513, 610480, 610516, 610521, 610419, 610415, 610416, 610456, 610461, 610460, 610472, 610470, 610477, 610471, 610463, 610462, 610467, 610469, 610465, 610587, 610639, 610614, 610616, 610608, 610605, 610645, 610634, 610623, 610626, 610615, 610633, 610622, 610609, 610621, 610672, 610678, 610674, 610671, 610667, 610673, 610675, 610677, 610650, 610662, 610692, 610646, 610693, 610652, 610661, 610663, 610654, 610655, 610689, 610653, 610656, 610657, 610694, 610688, 610658, 610660, 610725, 610729, 610732, 610731, 610719, 610726, 610703, 610709, 610718, 610715, 610717, 610716, 610698, 610702, 610714, 610711, 610705, 610706, 610697, 610710, 610755, 610757, 610754, 610756, 610767, 610766, 610769, 610768, 610785, 610788, 610800, 610793, 610794, 610795, 610801, 610814, 610811, 610758, 610786, 610744, 610804, 610736, 610737, 610742, 610750, 610802, 610806, 610791, 610759, 610762, 610738, 610739, 610740, 610741, 610792, 610805, 610808, 610733, 610761, 610763, 610798, 610803, 610735, 610810, 610760, 610749, 610799, 610809, 610764, 610745, 610743, 610746, 610829, 610832, 610833, 610830, 610831, 610827, 610828, 610825, 610826, 610822, 610807, 610813, 610821. UDI/DI 16954701000412 (Suzhou), Serial Numbers: 630104, 630105, 630160, 630159, 630162, 630110, 630140, 630175, 630143, 630103, 630108, 630174, 630099, 630112, 630113, 630136, 630107, 630102, 630100, 630106, 630115, 630114, 630132, 630111, 630101, 630109, 630146, 630139, 630116, 630126, 630117, 630118, 630120, 630119, 630121, 630122, 630123, 630124, 630125, 630129, 630128, 630130, 630131, 630134, 630135, 630137, 630141, 630142, 630133, 630127, 630164, 630172, 630173, 630144, 630156, 630145, 630157, 630158, 630155, 630154, 630149, 630152, 630170, 630171, 630150, 630151, 630153, 630163, 630148, 630147, 630138, 630194, 630197, 630202, 630203, 630178, 630184, 630166, 630186, 630182, 630176, 630205, 630179, 630180, 630181, 630195, 630187, 630161, 630169, 630167, 630165, 630193, 630191, 630192, 630190, 630177, 630188, 630183, 630196, 630199, 630168, 630219, 630208, 630217, 630230, 630223, 630214, 630210, 630235, 630232, 630201, 630226, 630227, 630218, 630233, 630221, 630222, 630216, 630215, 630206, 630207, 630211, 630231, 630209, 630224, 630234, 630228, 630220, 630225, 630229, 630204, 630212, 630213, 630259, 630260, 630252, 630281, 630282, 630272, 630265, 630279, 630256, 630255, 630269, 630242, 630248, 630241, 630264, 630275, 630270, 630249, 630253, 630271, 630267, 630278, 630254, 630263, 630243, 630246, 630273, 630280, 630238, 630251, 630257, 630258, 630274, 630240, 630244, 630247, 630262, 630237, 630266, 630236, 630261, 630245, 630307, 630300, 630301, 630315, 630284, 630319, 630327, 630296, 630292, 630304, 630305, 630311, 630310, 630312, 630324, 630291, 630321, 630323, 630325, 630289, 630316, 630317, 630308, 630306, 630318, 630299, 630314, 630302, 630313, 630290, 630285, 630298, 630293, 630295, 630303, 630320, 630309, 630294, 630287, 630286, 630331, 630288, 630322, 630328, 630337, 630351, 630341, 630345, 630344, 630343, 630342, 630340, 630334, 630330, 630332, 630336, 630338, 630333, 630339, 630347, 630346, 630348, 630326, 630198, 630189, 630276, 630353, 630185, 630352, 630329, 630277, 630283, 630297, 630349, 630200, 630239.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0429-2026
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0428-2026Class IIOngoingVoluntary: Firm initiated
Z-0429-2026Class IIOngoingVoluntary: Firm initiated