Z-0481-2026Class II

SimplyGo Mini, Standard Battery Kit

Respironics, Inc. / initiated 2025-10-07 / Open, Classified / root cause: manufacturing process

Philips Respironics, Inc. began a recall of SimplyGo Mini, Standard Battery Kit; Product code: 1116816; The affected battery kits are sold as accessories for the SimplyGo Mini: SimplyGo Mini Portable... on . FDA classified it as Class II and gives the reason as "Certain serial numbers of the SimplyGo Mini standard battery kit contain a defective chip which can lead the battery to either fail to charge or lose voltage output when in use in the device. If the fault occurs while...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0481-2026
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Respironics, Inc.
Product
SimplyGo Mini, Standard Battery Kit; Product code: 1116816; The affected battery kits are sold as accessories for the SimplyGo Mini: SimplyGo Mini Portable Oxygen Concentrator Product code: 1113600
Product code
CAW Generator, Oxygen, Portable
Reason for recall
Certain serial numbers of the SimplyGo Mini standard battery kit contain a defective chip which can lead the battery to either fail to charge or lose voltage output when in use in the device. If the fault occurs while the device is running on battery power, the SimplyGo Mini will shut down and stop providing oxygen to the user. This may happen at a temperature within the specified operating temperature range of the SimplyGo Mini.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
On October 7, 2025 URGENT MEDICAL DEVICE RECALL letters were sent to customers. Actions for Distributors: Check inventory and place into quarantine any identified affected products. Complete and return the provided response form. Alert any customers who may have purchased affected products and work with them to return the affected product. A Philips representative will reach out regarding return and replacement of all affected product. Actions for patients: Check the serial numbers to verify if product affected by recall. If your product is affected, please discontinue use or the battery immediately. Contact your distributor and follow their instructions on how to receive a replacement battery and arrange for the return of the affected batteries in your possession. If the serial number on your battery kit does not match any of the affected serial numbers, then your battery is not affected by this issue and there is no need for any further action by you.
Quantity in commerce
56 units (55 US, 1 OUS)
Distribution
US Nationwide distribution in the states of GA, ID, KY, NJ, NY and the country of Canada.
Product codes and lots
Product Code: 1116816; UDI-DI: 00606959032385; Serial Numbers: 2438V115830, 2438V115960, 2438V116033, 2438V115899, 2438V115832, 2438V115964, 2438V116034, 2438V115837, 2438V115965, 2438V116040, 2438V115840, 2438V115967, 2438V116044, 2438V115842, 2438V115968, 2438V116045, 2438V115844, 2438V115970, 2438V116046, 2438V115878, 2438V115972, 2438V116047, 2438V115890, 2438V115973, 2438V116048, 2438V115894, 2438V115974, 2438V116049, 2438V115895, 2438V115980, 2438V116050, 2438V115896, 2438V115981, 2438V116051, 2438V115897, 2438V115983, 2438V116052, 2438V115919, 2438V115984, 2438V116226, 2438V115937, 2438V115991, 2438V116229, 2438V115939, 2438V115992, 2438V116233, 2438V115942, 2438V115993, 2438V116249, 2438V115948, 2438V115994, 2438V116253, 2438V115957, 2438V115995, 2438V116258, 2438V117762;

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0481-2026
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0481-2026Class IIOngoingVoluntary: Firm initiated