Thermo scientific MAS Omni CARDIO, Liquid Assayed Integrated Cardiac Control, Model/Catalog Number: OCRD-L...

Microgenics Corporation / initiated 2025-10-01 / Open, Classified / root cause: manufacturing process

Microgenics Corporation began a recall of Thermo scientific MAS Omni CARDIO, Liquid Assayed Integrated Cardiac Control, Model/Catalog Number: OCRD-L, OCRD-101, OCRD-202, OCRD-303, OCRD-SP, OCRD-MP... on . FDA classified it as Class II and gives the reason as "The firm received complaints from customers reporting vial-to-vial variability when using the control with high sensitivity Troponin I assays. This defect may lead to as delay in patient results for hs Troponin I assays". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0415-2026
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Microgenics Corporation
Product
Thermo scientific MAS Omni CARDIO, Liquid Assayed Integrated Cardiac Control, Model/Catalog Number: OCRD-L, OCRD-101, OCRD-202, OCRD-303, OCRD-SP, OCRD-MP, OCRD-UL; Product Description: In Vitro Diagnostic external control
Product code
JJY Multi-Analyte Controls, All Kinds (Assayed)
Reason for recall
The firm received complaints from customers reporting vial-to-vial variability when using the control with high sensitivity Troponin I assays. This defect may lead to as delay in patient results for hs Troponin I assays.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
ThermoFisher Scientific issued an URGENT -MEDICAL DEVICE RECALL notice to its consignees on 10/02/2025 via email. The notice explained the issue, risk to health, and requested the following: "ACTIONS TO BE TAKEN BY A USER 1. If you are a user of the High sensitivity Troponin I Assay, determine if you have inventory of the affected lots of MAS Omni"CARDIO" control as listed in Table 1. As stated in the instructions for use, instrument values provided are specific to this lot of control only and are intended to assist the laboratory in establishing its own means and ranges. 2. You may continue to use controls that meet your laboratory s performance requirements. 3. If the variability of the material exceeds your laboratory s specification, dispose of the material in accordance with local, regional, and national regulations. After disposing of the material, contact [email protected] for a credit. 4. Customers using the affected product with standard/ regular Troponin I assays, can continue to use the product as long as it meets their laboratory s specification and continue monitoring performance of Troponin I in MAS Omni"CARDIO" control. 5. This recall notice needs to be passed to anyone within your organization or any organization where the potentially affected lots have been transferred. 6. Retain a copy of this letter for your laboratory records. 7. Please complete the Recall Response Form attached to this Notification within 10 days of the date of this letter and return to Thermo Fisher as instructed in the form to E-mail:[email protected] 8. If you have any additional questions regarding this Recall Notice, contact Thermo Fisher Scientific Technical Support at toll free 1-800-232-3342 (USA and Canada only) or 510-979-5000 (International), Option 2, then option 3 or by email: [email protected]." "ACTIONS TO BE TAKEN BY A DISTRIBUTOR 1. Please notify your customers of this Recall Notice using this recall notice and request t
Quantity in commerce
4462
Distribution
Worldwide - US Nationwide distribution in the states of AL, OK, NH, NY, WI, IL, LA, AR, TX, UT, GA, ID and the countries of MEX, CAN, KOR, MYS, DEU, GBR, BRA, JAP, HKG, NZL, AUS.
Product codes and lots
OCRD-UL: UDI/DI 00884883011946, Lot Number OCRD2704U, Exp 2027-04-30; OCRD-L: UDI/DI 00884883011953, Lot Number OCRD2704L, Exp 2027-04-30; OCRD-101: UDI/DI 00884883011960, Lot Number OCRD27041, Exp 2027-04-30; OCRD-202: UDI/DI 00884883011977, Lot Number OCRD27042, Exp 2027-04-30; OCRD-303: UDI/DI 00884883011984, Lot Number OCRD27043, Exp 2027-04-30; OCRD-SP: UDI/DI 00884883011991, Lot Number OCRD2704S, Exp 2027-04-30; OCRD-MP: UDI/DI 00884883012004, Lot Number OCRD2704M, Exp 2027-04-30.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0415-2026
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0415-2026Class IIOngoingVoluntary: Firm initiated