Z-0478-2026Class II

Virata Preassembled Polyaxial Screw, utilized in the Virata Spinal Fixation System. The following sizes are...

Orthofix SRL / initiated 2025-09-19 / Open, Classified

Orthofix U.S. LLC began a recall of Virata Preassembled Polyaxial Screw, utilized in the Virata Spinal Fixation System. The following sizes are affected: 1. POLYAXIAL SCREW, CC, 5.5mm X 30mm... on . FDA classified it as Class II and gives the reason as "Polyaxial screws do not meet performance standards". As of , the recall status on record is Open, Classified.

Recall record

Recall number
Z-0478-2026
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Orthofix SRL
Product
Virata Preassembled Polyaxial Screw, utilized in the Virata Spinal Fixation System. The following sizes are affected: 1. POLYAXIAL SCREW, CC, 5.5mm X 30mm; Item Number: TL1-10755030. 2. POLYAXIAL SCREW, CC, 5.5mm X 35mm; Item Number: TL1-10755035. 3. POLYAXIAL SCREW, CC, 5.5mm X 40mm; Item Number: TL1-10755040. 4. POLYAXIAL SCREW, CC, 5.5mm X 45mm; Item Number: TL1-10755045. 5. POLYAXIAL SCREW, CC, 5.5mm X 50mm; Item Number: TL1-10755050. 6. POLYAXIAL SCREW, CC, 6.5mm X 30mm; Item Number: TL1-10765030. 7. POLYAXIAL SCREW, CC, 6.5mm X 35mm; Item Number: TL1-10765035. 8. POLYAXIAL SCREW, CC, 6.5mm X 40mm; Item Number: TL1-10765040. 9. POLYAXIAL SCREW, CC, 6.5mm X 45mm; Item Number: TL1-10765045. 10. POLYAXIAL SCREW, CC, 6.5mm X 50mm; Item Number: TL1-10765050. 11. POLYAXIAL SCREW, CC, 6.5mm X 55mm; Item Number: TL1-10765055. 12. POLYAXIAL SCREW, CC, 6.5mm X 60mm; Item Number: TL1-10765060. 13. POLYAXIAL SCREW, CC, 7.5mm X 30mm; Item Number: TL1-10775030. 14. POLYAXIAL SCREW, CC, 7.5mm X 35mm; Item Number: TL1-10775035. 15. POLYAXIAL SCREW, CC, 7.5mm X 40mm; Item Number: TL1-10775040. 16. POLYAXIAL SCREW, CC, 7.5mm X 45mm; Item Number: TL1-10775045. 17. POLYAXIAL SCREW, CC, 7.5mm X 50mm; Item Number: TL1-10775050. 18. POLYAXIAL SCREW, CC, 7.5mm X 55mm; Item Number: TL1-10775055. 19. POLYAXIAL SCREW, CC, 7.5mm X 60mm; Item Number: TL1-10775060. 20. POLYAXIAL SCREW, CC, 8.5mm X 35mm; Item Number: TL1-10785035. 21. POLYAXIAL SCREW, CC, 8.5mm X 40mm; Item Number: TL1-10785040. 22. POLYAXIAL SCREW, CC, 8.5mm X 45mm; Item Number: TL1-10785045. 23. POLYAXIAL SCREW, CC, 8.5mm X 50mm; Item Number: TL1-10785050. 24. POLYAXIAL SCREW, CC, 8.5mm X 55mm; Item Number: TL1-10785055. 25. POLYAXIAL SCREW, CC, 8.5mm X 60mm; Item Number: TL1-10785060.
Product code
NKB Thoracolumbosacral Pedicle Screw System
Reason for recall
Polyaxial screws do not meet performance standards.
FDA root cause
Process change control
Action
Orthofix notified consignees via email on 09/19/2025. Consignees were instructed to immediately discontinue use of the affected devices, quarantine all affected inventory, and return units to Orthofix using the provided Return Goods Authorization (RGA) process. Surgeons were instructed to monitor affected patients according to standard clinical practice. Customers were advised not to implant or distribute any remaining affected product. All consignees were requested to complete and return the response form.
Quantity in commerce
624 units
Distribution
US Nationwide distribution in the states of California, North Carolina, and Texas.
Product codes and lots
1. Item Number: TL1-10755030. UDI-DI: 10889981335229. Lot Number: JP2572B. 2. Item Number: TL1-10755035. UDI-DI: 10889981336165. Lot Number: JP2757B. 3. Item Number: TL1-10755040. UDI-DI: 10889981335717. Lot Number: JP2758B. 4. Item Number: TL1-10755045. UDI-DI: 10889981335953. Lot Numbers: JP2759B, JP2799B, JP2800B, JP2880B. 5. Item Number: TL1-10755050. UDI-DI: 10889981336318. Lot Numbers: JP2760B, JP2801B. 6. Item Number: TL1-10765030. UDI-DI: 10889981335885. Lot Number: JP2655B. 7. Item Number: TL1-10765035. UDI-DI: 10889981335540. Lot Number: JP2656B. 8. Item Number: TL1-10765040. UDI-DI: 10889981336332. Lot Number: JP2657B, JP2791B. 9. Item Number: TL1-10765045. UDI-DI: 10889981335526. Lot Numbers: JP2658B, JP2659B, JP2792B. 10. Item Number: TL1-10765050. UDI-DI: 10889981335342. Lot Numbers: JP2660B, JP2794B, JP2795B. 11. Item Number: TL1-10765055. UDI-DI: 10889981335168. Lot Number: JP2661B. 12. Item Number: . UDI-DI: . Lot Number: Item Number: TL1-10765060. UDI-DI: 10889981335816. Lot Number: JP2662B. 13. Item Number: TL1-10775030. UDI-DI: 10889981335427. Lot Number: JP2663B. 14. Item Number: TL1-10775035. UDI-DI: 10889981335632. Lot Number: JP2668B. 15. Item Number: TL1-10775040. UDI-DI: 10889981335113. Lot Number: JP2669B. 16. Item Number: TL1-10775045. UDI-DI: 10889981335403. Lot Number: JP2670B. 17. Item Number: TL1-10775050. UDI-DI: 10889981335830. Lot Number: JP2671B. 18. Item Number: TL1-10775055. UDI-DI: 10889981335250. Lot Numbers: JP2672B, JP2887B. 19. Item Number: TL1-10775060. UDI-DI: 10889981335076. Lot Number: JP2673B. 20. Item Number: TL1-10785035. UDI-DI: 10889981335410. Lot Number: JP2674B. 21. Item Number: TL1-10785040. UDI-DI: 10889981335779. Lot Number: JP2675B. 22. Item Number: TL1-10785045. UDI-DI: 10889981335243. Lot Number: JP2664B. 23. Item Number: TL1-10785050. UDI-DI: 10889981335847. Lot Number: JP2665B. 24. Item Number: TL1-10785055. UDI-DI: 10889981335106. Lot Number: JP2666B. 25. Item Number: TL1-10785060. UDI-DI: 10889981334949. Lot Number: JP2667B.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0478-2026
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0478-2026Class IIOngoingVoluntary: Firm initiated