Z-0534-2026Class II
Brand Name: ReQuest Measles IgG Product Name: Rubeola IgG Enzyme Linked Immunoabsorbant Assay Model/Catalog...
Quest International, Inc. / initiated 2025-10-01 / Open, Classified
Quest International, Inc. began a recall of Brand Name: ReQuest Measles IgG Product Name: Rubeola IgG Enzyme Linked Immunoabsorbant Assay Model/Catalog Number: 01-190 Software Version: N/A Product... on . FDA classified it as Class II and gives the reason as "Due to distributing the measles IgG IVD without a premarket approved/cleared". As of , the recall status on record is Open, Classified.
Recall record
- Recall number
- Z-0534-2026
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Quest International, Inc.
- Product
- Brand Name: ReQuest Measles IgG Product Name: Rubeola IgG Enzyme Linked Immunoabsorbant Assay Model/Catalog Number: 01-190 Software Version: N/A Product Description: Measles IgG tests, performed using ELISA (Enzyme-Linked Immunosorbent Assay) method, used to detect the presence of IgG antibodies against measles virus in a patient's serum or plasma Component: not a component
- Product code
- LJB Enzyme Linked Immunoabsorbent Assay, Rubeola Igg
- Reason for recall
- Due to distributing the measles IgG IVD without a premarket approved/cleared.
- FDA root cause
- No Marketing Application
- Action
- On 10/01/2025, the firm email an "URGENT: Medical Device Recall Measles IgG IFU" letter to customers informing them that the Measles IgG Medical Device is subject to the limitations of exemption from premarket notification (510(k)) under 21 CFR866.9(c)(7) and is subject to the 510(k) premarket requirements prior to being marketed and distributed in the U.S. Customers are instructed to examine their inventory, discontinue use and quarantine the affected lot and review the updated Instruction for Use (IFU) and ensure all relevant personnel are informed of the changes. The previous version of the IFU should be replace with the updated version, Revision 9. The updated IFU (Revision 9) can also be requested by contacting Customer Support team at [email protected]. Awareness Technology, Inc. will be revising the intended use of the product to be only for qualitative detection of human IgG antibodies to measles virus in human serum by enzyme immunoassay. The company will also be removing language in the Instructions for Use (IFU) that could be interpreted as determining immune status, statements regarding paired sera for demonstrating seroconversion or significant rise in antibody levels, and any reference to semi-quantitative procedures. The updated IFU (Revision 9) will be included in all product shipments starting 10/01/2025. For questions or assistance, contact Customer Service at [email protected].
- Distribution
- All products were distributed in the United States. (Utah, Washington).
- Product codes and lots
- Lot Code: DI# 00850487007357 Lots affected: L07024 and E03025