Zimmer Biomet Dual Mobility Vivacit-E Highly Crosslinked Polyethylene, 28 mm I.D., 44 mm O.D., Size F...
Zimmer, Inc. began a recall of Zimmer Biomet Dual Mobility Vivacit-E Highly Crosslinked Polyethylene, 28 mm I.D., 44 mm O.D., Size F, Bearing, Model/Catalog Number: 110031012; semi... on . FDA classified it as Class II and gives the reason as "Zimmer Biomet is conducting a medical device recall for two (2) lots of the Dual Mobility Vivacit-E Bearing due to a commingle. The outer package is labeled as a Size F 44 mm, however, the implant inside the package is...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-0418-2026
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Zimmer GmbH
- Product
- Zimmer Biomet Dual Mobility Vivacit-E Highly Crosslinked Polyethylene, 28 mm I.D., 44 mm O.D., Size F, Bearing, Model/Catalog Number: 110031012; semi constrained hip prosthesis
- Product code
- LPH Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented
- Reason for recall
- Zimmer Biomet is conducting a medical device recall for two (2) lots of the Dual Mobility Vivacit-E Bearing due to a commingle. The outer package is labeled as a Size F 44 mm, however, the implant inside the package is a Size G 46 mm, and vice versa.
- Root cause tag
- labeling and UDI
- FDA root cause
- Labeling mix-ups
- Action
- Zimmer Biomet issued an URGENT MEDICAL DEVICE recall notice to its consignees on 9/25/2025 via FedEx, 2nd Day Delivery. The notice explained the issue, potential risks, and requested the return of all affected devices in inventory. Risk Managers were notified that the sales representatives would be removing the affected products. Recommendation for patient monitoring was made to the implanting surgeons. For further questions or concerns after reviewing this notice, please call customer service at 574-371-3071 between 8:00 a.m. and 5:00 p.m. EST, Monday through Friday. Calls received outside of call center operating hours will receive a voicemail prompt or be transferred to an on-call representative in the event of an emergency. Alternatively, questions may be emailed to [email protected].
- Quantity in commerce
- 72 units
- Distribution
- US: OH, GA, WI, NE, IA, IL, CA, NY, MD, NC, PA, AL, TN, MI, WA, FL, TX, KS, NM, AZ, and CANADA, AUSTRALIA, NETHERLANDS
- Product codes and lots
- UDI (01)00889024572706(17)300406(10)67160447; Lot Number 67160447
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0418-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0419-2026 | Class II | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.