Multichem ID-B. Model/Catalog Number: SR103B Intended for use as an unassayed positive qualitative quality...
TECHNO-PATH MANUFACTURING LTD. began a recall of Multichem ID-B. Model/Catalog Number: SR103B Intended for use as an unassayed positive qualitative quality control serum to monitor the precision of laboratory... on . FDA classified it as Class II and gives the reason as "Potential of negative recovery of the HBsAg analyte across two (2) lots of Multichem ID-B Positive Control when tested on the Alinity i platform". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0501-2026
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Techno-Path Manufacturing Ltd.
- Product
- Multichem ID-B. Model/Catalog Number: SR103B Intended for use as an unassayed positive qualitative quality control serum to monitor the precision of laboratory testing procedures for the determination of anti-HIV-1 IgG, anti-HBc IgG, anti-HTLV-I IgG, anti-HCV IgG, HBs antigen and anti-Treponema pallidum IgG.
- Product code
- OHQ Multi-Analyte Controls Unassayed
- Reason for recall
- Potential of negative recovery of the HBsAg analyte across two (2) lots of Multichem ID-B Positive Control when tested on the Alinity i platform.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- LGC Clinical Diagnostics notified consignees on about 10/17/2025 via letter that Techno-path Manufacturing Ltd was recalling specific Multichem ID-B products. Consignee were informed that the Instructions for Use (IFU) for Multichem ID-B SR103B, Lots SB140524P and SB161024P have been updated to remove the representative reactivity for the analyte HBsAg; therefore, no representative reactivity result will be provided for HBsAg in these lots. Consignees were instructed to discard all previous versions of the IFU for affected lots, share the information with all applicable laboratory staff or customers if further distributed, retain the notification as part of the laboratory Quality System documentation, and complete and return the provided response form. Techno-path Manufacturing Ltd. is investigating this issue and will implement a correction for all future lots of Multichem ID-B Positive Control SR103B that will be released.
- Quantity in commerce
- 76 units
- Distribution
- US distribution to CA, CO, CT, FL, HI, IL, LA, MI, MN, NV, NY, OR, SC, TN, TX, WI, WY.
- Product codes and lots
- Model Number: SR103B. UDI: (01)05391523442312(240)SR103B(17)260531(10)SB140524P. Lot number: SB140524P. Expiration Date: 31-5-2026
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0501-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0502-2026 | Class II | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.