Z-0502-2026Class II

Multichem ID-B. Model Number: SR103B. Intended for use as an unassayed positive qualitative quality control...

Techno-Path Manufacturing Ltd. / initiated 2025-10-17 / Open, Classified / root cause: manufacturing process

TECHNO-PATH MANUFACTURING LTD. began a recall of Multichem ID-B. Model Number: SR103B. Intended for use as an unassayed positive qualitative quality control serum to monitor the precision of laboratory... on . FDA classified it as Class II and gives the reason as "Potential of negative recovery of the HBsAg analyte across two (2) lots of Multichem ID-B Positive Control when tested on the Alinity i platform". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0502-2026
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Techno-Path Manufacturing Ltd.
Product
Multichem ID-B. Model Number: SR103B. Intended for use as an unassayed positive qualitative quality control serum to monitor the precision of laboratory testing procedures for the determination of anti-HIV-1 IgG, anti-HBc IgG, anti-HTLV-I IgG, anti-HCV IgG, HBs antigen and anti-Treponema pallidum IgG.
Product code
OHQ Multi-Analyte Controls Unassayed
Reason for recall
Potential of negative recovery of the HBsAg analyte across two (2) lots of Multichem ID-B Positive Control when tested on the Alinity i platform.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
LGC Clinical Diagnostics notified consignees on about 10/17/2025 via letter that Techno-path Manufacturing Ltd was recalling specific Multichem ID-B products. Consignee were informed that the Instructions for Use (IFU) for Multichem ID-B SR103B, Lots SB140524P and SB161024P have been updated to remove the representative reactivity for the analyte HBsAg; therefore, no representative reactivity result will be provided for HBsAg in these lots. Consignees were instructed to discard all previous versions of the IFU for affected lots, share the information with all applicable laboratory staff or customers if further distributed, retain the notification as part of the laboratory Quality System documentation, and complete and return the provided response form. Techno-path Manufacturing Ltd. is investigating this issue and will implement a correction for all future lots of Multichem ID-B Positive Control SR103B that will be released.
Quantity in commerce
51 units
Distribution
US distribution to CA, CO, CT, FL, HI, IL, LA, MI, MN, NV, NY, OR, SC, TN, TX, WI, WY.
Product codes and lots
Model Number: SR103B. UDI: (01)05391523442312(240)SR103B(17)261031(10) SB161024P. Lot Number: SB161024P. Expiration Date: 31-10-2026.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0502-2026
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0501-2026Class IIOngoingVoluntary: Firm initiated
Z-0502-2026Class IIOngoingVoluntary: Firm initiated