Z-0509-2026Class II

MC3 VitalFlow Console, REF 58100; Blood pump of ecmo

Medtronic Perfusion Systems / initiated 2025-10-03 / Open, Classified / root cause: software

Medtronic Perfusion Systems began a recall of MC3 VitalFlow Console, REF 58100; Blood pump of ecmo on . FDA classified it as Class II and gives the reason as "As of August 6, 2025, Medtronic has received eleven reports of VitalFlow Consoles displaying an E70 error code during normal operation. When this occurs, the touch screen may become temporarily unresponsive and go blank...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.

Recall record

Recall number
Z-0509-2026
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Medtronic Perfusion Systems
Product
MC3 VitalFlow Console, REF 58100; Blood pump of ecmo
Product code
QNR Blood Pump For Ecmo, Long-Term (> 6 Hours) Use
Reason for recall
As of August 6, 2025, Medtronic has received eleven reports of VitalFlow Consoles displaying an E70 error code during normal operation. When this occurs, the touch screen may become temporarily unresponsive and go blank for up to two minutes before recovering to full functionality. Importantly, the console continues to maintain set pump speed and function throughout.
Root cause tag
software
FDA root cause
Software in the Use Environment
Action
Medtronic Cardiac Surgery issued an Urgent: Medical Device Correction notice to its consignees on 10/03/2025. The notice explained the issue, potential harms, and requested the consignee contact Medtronic Sales Representative to schedule the service event to their affected VitalFlow" Console. Please reference Technical Service Update (TSU) "MOD0209" when scheduling. The notice also indicated that the consignees use unaffected VitalFlow Console devices, if available, until the correction has been completed. For questions regarding the communication, please contact your Medtronic Sales Representative.
Quantity in commerce
167 units
Distribution
US Nationwide distribution.
Product codes and lots
UDI/DI 10850007732247, Serial Numbers: 000023, 000047, 000073, 000100, 000126, 000150, 000173, 000197, 000223, 000024, 000048, 000075, 000101, 000127, 000151, 000174, 000202, 000224, 000025, 000051, 000076, 000102, 000128, 000152, 000177, 000203, 000225, 000027, 000052, 000078, 000103, 000129, 000153, 000179, 000204, 000228, 000028, 000054, 000079, 000104, 000130, 000154, 000180, 000205, 000229, 000029, 000055, 000080, 000106, 000131, 000155, 000181, 000206, 000277, 000030, 000056, 000081, 000108, 000132, 000156, 000182, 000207, 000284, 000031, 000058, 000082, 000109, 000133, 000157, 000183, 000208, 000032, 000059, 000083, 000110, 000134, 000159, 000184, 000209, 000033, 000060, 000084, 000112, 000135, 000161, 000185, 000210, 000034, 000061, 000085, 000115, 000136, 000162, 000186, 000211, 000035, 000063, 000087, 000116, 000140, 000163, 000187, 000212, 000037, 000065, 000088, 000117, 000141, 000165, 000188, 000213, 000038, 000066, 000089, 000118, 000142, 000166, 000189, 000214, 000039, 000067, 000090, 000119, 000143, 000167, 000190, 000215, 000040, 000068, 000092, 000120, 000145, 000168, 000191, 000216, 000043, 000069, 000095, 000121, 000146, 000169, 000192, 000217, 000044, 000070, 000096, 000122, 000147, 000170, 000193, 000218, 000045, 000071, 000097, 000124, 000148, 000171, 000194, 000219, 000046, 000072, 000099, 000125, 000149, 000172, 000195, 000221.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0509-2026
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0509-2026Class IIOngoingVoluntary: Firm initiated