IMMULITE 2000 Oak Mix Specific Allergen; Siemens Material Number (SMN): 10386105; Catalog Number: L2KUN6
Siemens Healthcare Diagnostics, Inc. began a recall of IMMULITE 2000 Oak Mix Specific Allergen; Siemens Material Number (SMN): 10386105; Catalog Number: L2KUN6 on . FDA classified it as Class II and gives the reason as "The barcode orientation on affected devices causes incorrect scanning order of tubes within the allergen holder wedge. When using an IMMULITE 2000 / IMMULITE 2000 XPi instrument in combination with a Zebra barcode...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-0491-2026
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Product
- IMMULITE 2000 Oak Mix Specific Allergen; Siemens Material Number (SMN): 10386105; Catalog Number: L2KUN6;
- Product code
- DHB System, Test, Radioallergosorbent (Rast) Immunological
- Reason for recall
- The barcode orientation on affected devices causes incorrect scanning order of tubes within the allergen holder wedge. When using an IMMULITE 2000 / IMMULITE 2000 XPi instrument in combination with a Zebra barcode scanner, the instrument will not correctly recognize the position of tubes within the allergen holder wedge.
- Root cause tag
- labeling and UDI
- FDA root cause
- Under Investigation by firm
- Action
- On October 6, 2025 URGENT MEDICAL DEVICE CORRECTION letters were sent to customers. Customer Actions Perform the instructions provided below: " Please review this letter with your Medical Director to determine the appropriate course of action, including for any previously generated results, if applicable. " Discontinue use of and discard IMMULITE 2000 Oak Mix Specific Allergen Lot 627. " Review your inventory of the product to determine your laboratory s replacement needs and to provide information to Siemens Healthineers for reporting to the authorities. " Complete and return the Field Correction Effectiveness Check Form attached to this letter within 7 days. " Please retain this letter with your laboratory records and forward this letter to those who may have received this product. Resolution The issue is resolved with the next available lot of IMMULITE 2000 Oak Mix Specific Allergen. It is anticipated that the replacement lot will be available by mid-November 2025. We apologize for the inconvenience this situation may cause. If you have any questions, please contact your Siemens Healthineers Custom
- Quantity in commerce
- 34 units (1 US, 33 OUS)
- Distribution
- Worldwide - US Nationwide distribution in the state of CA and the countries of Argentina, Australia, Austria, Belgium, Bosnia Herzeg. , Brazil, Canary Islands, Chile, Colombia, Costa Rica, Croatia, Czech Republic, Estonia, France, Germany, Greece, Guatemala, Hong Kong, Hungary, Israel, Italy, Latvia, Macedonia, Mexico, Netherlands, Norway, Pakistan, Panama, Poland, Portugal, Romania, Russian Fed., Serbia, Slovakia, Spain, Sweden, Turkey, U.A.E., Ukraine, United Kingdom, Uruguay.
- Product codes and lots
- Siemens Material Number (SMN): 10386105; Model/Catalog: T77L2; UDI-DI: 00630414965970; Lot number: 627;
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0491-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0492-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0493-2026 | Class II | Ongoing | Voluntary: Firm initiated |