bk3000 Ultrasound System w/battery; Model No. 2300-66; System, Imaging, Pulsed Doppler, Ultrasonic
B-K Medical A/S began a recall of bk3000 Ultrasound System w/battery; Model No. 2300-66; System, Imaging, Pulsed Doppler, Ultrasonic on . FDA classified it as Class II and gives the reason as "GE HealthCare has become aware that the Instructions for Use within the user guide and service manual do not adequately state the required battery replacement intervals, end-of-life handling procedures, or general...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0438-2026
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- BK Medical ApS
- Product
- bk3000 Ultrasound System w/battery; Model No. 2300-66; System, Imaging, Pulsed Doppler, Ultrasonic
- Product code
- IYN System, Imaging, Pulsed Doppler, Ultrasonic
- Reason for recall
- GE HealthCare has become aware that the Instructions for Use within the user guide and service manual do not adequately state the required battery replacement intervals, end-of-life handling procedures, or general battery safety precautions.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- GE Healthcare issued an "URGENT MEDICAL DEVICE CORRECTION NOTICE" to its consignees on 10/09/2025 via traceable means. The notice explained the issue, potential risk, and requested: "Actions to be taken by Customer / User You can continue to use your device by following the instructions below: 1. Review and follow the applicable addendums to the Instructions for Use and the updated Service Manuals. The Addendums and Service Manuals are available for download at: https://www.gehealthcare.com/support/manuals. 2. Follow the conditions defined within the addendum and Service Manual applicable to your product for further instructions. 3. Place the applicable addendum and Service Manual with your existing product Instructions for Use. 4. If you have difficulty accessing the documentation or require a printed copy of the applicable addendum or Service Manual, please contact a GE HealthCare Service Representative for assistance. Please ensure all potential users in your facility are made aware of this safety notification and the required actions. Please retain this document for your records." For questions or concerns regarding this notification, please contact GE HealthCare at 1-800-645-4442 or [email protected] or your local Service Representative.
- Quantity in commerce
- 496 units
- Distribution
- Worldwide distribution: US (nationwide) and OUS (foreign) to countries of: Algeria, Argentina, Australia, Austria, Belgium, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Czech Republic, Denmark, Dominican Rep, Ecuador, Estonia, Finland, France, Georgia, Germany, Greece, Hong Kong, Hungary, Iceland, India, Indonesia, Ireland, Israel, Italy, Japan, Korea (Republic Of), Kuwait, Latvia, Lebanon, Luxembourg, Malaysia, Mexico, Mongolia, Morocco, Nepal, Netherlands, New Zealand, Norway, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkiye, Ukraine, United Arab Emirates, and United Kingdom.
- Product codes and lots
- GTIN 05704916000264 and pending, Serial Numbers: 5016428, 5016429, 5016430, 2009154, 2008179, 2004456, 2007484, 2005976, 2007361, 2009167, 2003895, 2004700, 2008564, 2008948, 2004560, 2008452, 2008423, 2005567, 2006788, 2002815, 2008537, 2008687, 2003294, 2005833, 2008560, 2005374, 2008151, 2008871, 2008873, 2007476, 2002562, 2005398, 2005950, 2008456, 2004746, 2004957, 2006757, 2001059, 5016374, 2006796, 2001364, 2004636, 2008856, 2008935, 2002361, 2005934, 2008793, 2008194, 2003909, 2006125, 2008230, 2008254, 2008489, 2008553, 5016508, 2005822, 2009176, 2001734, 2007607, 2007483, 2001065, 2007453, 2007710, 2004474, 2004670, 2004737, 2008881, 2004632, 2005554, 2007632, 2003894, 2005918, 2006074, 2006682, 2006840, 5016482, 2005845, 2006294, 2006298, 2007485, 2005580, 5016272, 2001735, 2003293, 2006659, 2003287, 2007692, 2007725, 2007693, 2008451, 2008886, 2007724, 2003890, 2001729, 2005522, 2006364, 2007711, 2008471, 2004725, 2004735, 2007491, 2008182, 2002118, 5016451, 2007325, 2007363, 2007454, 2007455, 2005564, 2005798, 2008914, 2005797, 2005821, 5016505, 5016509, 2007650, 2008218, 2001728, 2008159, 2003288, 2001366, 2004364, 2004368, 2009156, 2008146, 2008210, 2008530, 2004363, 2002120, 2000761, 2003776, 2008522, 2008892, 2002001, 2004482, 2005382, 2003170, 2003171, 2005191, 2007477, 2005303, 2002117, 2005973, 2005220, 2007356, 2008562, 2007649, 2005951, 2004686, 2008481, 2005201, 2003896, 2005903, 2005912, 2005947, 2005966, 2006008, 2001999, 2005882, 2004962, 2002565, 2008167, 2008900, 2001359, 2005413, 2006720, 2008684, 2008925, 2006604, 2004387, 2008864, 2008874, 2002114, 2002559, 2003942, 2009173, 2008248, 2008266, 2007653, 2005171, 2008450, 2001057, 2001064, 5016507, 2006006, 2004652, 2007492, 2001363, 2008497, 2001737, 2005298, 2008857, 2001063, 2005919, 2007359, 5016415, 2003167, 2006673, 5016416, 2008866, 5016506, 2008963, 2005543, 2007684, 2006826, 2004695, 2008880, 2008251, 2009256, 2002556, 2002563, 2002115, 2006773, 2007761, 2006857, 2008243, 2008268, 2003174, 2006308, 2006605, 2005968, 2008227, 2005390, 2008854, 2002003, 2002005, 2008453, 2005275, 2008508, 2007481, 2002557, 2005801, 2002007, 2002000, 2005192, 2004459, 2004559, 2004654, 2004712, 2004720, 2008202, 2008868, 2005225, 2000765, 2002002, 2008927, 2006057, 2002116, 2003962, 2005898, 2006493, 2007683, 2005294, 2005317, 2005254, 2004692, 2004696, 2004365, 2004447, 2004789, 2002006, 2006078, 2009263, 2007360, 2007498, 2001361, 2001367, 2003168, 2007631, 2008550, 2008937, 2005265, 2002819, 2001362, 2003295, 2005386, 2003883, 2003957, 2004409, 2004627, 2003906, 2007355, 2004450, 2006080, 2005924, 2002119, 2002121, 2005373, 2003166, 2001360, 2006527, 2004452, 2002816, 2002818, 2001730, 2006127, 2001998, 2008689, 2006063, 2008241, 2007451, 2006096, 2006296, 2006358, 2005278, 2007609, 2007610, 2007497, 2008878, 2008953, 2007478, 2007702, 2005379, 2008606, 2001731, 2004394, 2008579, 2002122, 2004485, 2001732, 2007654, 2005926, 2005354, 2005954, 2005925, 2005187, 5016341, 2006777, 2004527, 2004448, 2004451, 2006092, 2002004, 2007354, 2001066, 2005975, 2007463, 2001736, 2004667, 2004685, 2002820, 2003893, 2001060, 2000763, 2005974, 2002558, 2002817, 2004633, 2005955, 2004529, 2004415, 2005405, 2007452, 2005540, 2005168, 2000762, 2003859, 2004366, 2004390, 2004477, 2004713, 2004748, 2005215, 2005304, 2007328, 2004749, 2001365, 2005308, 2005365, 2004541, 2004719, 2006833, 2007531, 2005593, 2003175, 2003173, 2005579, 2005719, 2000764, 2001058, 2003292, 2004054, 2006134, 2008169, 2008233, 2008271, 2008323, 2008325, 2008535, 2008556, 2008568, 2008870, 2008872, 2008960, 2009188, 2008958, 2004053, 2006095, 2001062, 2005313, 2005395, 2007701, 2007326, 2004752, 2007353, 2004539, 2008237, 2008421, 2008797, 2005942, 2006603, 2007482, 2007629, 2007734, 2002564, 2007322, 2005902, 2008954, 2004738, 2003286, 2004449, 2007357, 2007358, 2006061, 2008859, 2005742, 2008921, 2002264, 2004732, 2002113, 2002814, 2004642, 2007636, 2005295, 2005389, 2007461, 2008141, 2008327, 2008791, 2008801, 2008883, 2008916, 2008682, 2008861, 2008934, 2008955, 2008950, 2004745, 2006282, 2006481, 2006784, 2005305, 2001061, 2001368, 2001733, 2005388, 2006790, 2008515, 2006665, 2008513, 2007635, 2004731, 2006699, 2005383, 2003963, 2005849, 2004012, 2006076, 2005933, 2006059, 2006345, 2006373, 2005401, 2003777, 2003881, 2003862, 2004462, 2004407, 2003959, 2004664, 2006093, 2006128, 2003172, 2003917, 2005410, 2005375, 2005802, 2003897, 2003958, 2007462, 2008155, 2008163, 2008173, 2008258, 2004389, 2004684.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0435-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0436-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0437-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0438-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0439-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0440-2026 | Class II | Ongoing | Voluntary: Firm initiated |