Quantum Intersegmental Table, Model/Catalog Number: Q400; heated massage table
Pivotal Health Solutions, Inc. began a recall of Quantum Intersegmental Table, Model/Catalog Number: Q400; heated massage table on . FDA classified it as Class II and gives the reason as "The heat function may become damaged over time due to patients pressing on the areas with heat function components. Damage to components can lead to overheating which can lead to smoldering and melting of the vinyl...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0587-2026
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Pivotal Health Solutions
- Product
- Quantum Intersegmental Table, Model/Catalog Number: Q400; heated massage table
- Product code
- ISA Massager, Therapeutic, Electric
- Reason for recall
- The heat function may become damaged over time due to patients pressing on the areas with heat function components. Damage to components can lead to overheating which can lead to smoldering and melting of the vinyl cushion.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Pivotal Health Solutions issued an URGENT: MEDICAL DEVICE RECALL notice to its consignees on 10/9/2025 via certified mail and Fed Ex. The notice explained the issue with the device, risk to health and requested the following: Immediately stop using table(s) with heat. The heat function must be disabled. You have been identified as having been sold a Quantum Table with heat. The serial numbers of the tables impacted are below. If you are unsure if your table has a heat function, please identify the serial number on the Quantum table(s), located on the lower section of table by the power cable inlet, and verify the serial number is on the list. The notice also supplied instructions for disconnecting and disabling the heat function o the Q400 tables. Contact information for questions: o Matt White ([email protected]) or Julee Driver ([email protected]), Quality Dept, at 605-753-0110 o Brenda Anderson ([email protected]), Service Dept at 605-753-0110
- Quantity in commerce
- 34 units
- Distribution
- Distribution to US nationwide, Canada and Japan.
- Product codes and lots
- All serial numbers.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0587-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0588-2026 | Class II | Ongoing | Voluntary: Firm initiated |