CareLink Clinic, REF: MMT-7350
Medtronic MiniMed, Inc. began a recall of CareLink Clinic, REF: MMT-7350 on . FDA classified it as Class II and gives the reason as "Software error causing incorrect data to be displayed on the 24-hour Sensor Glucose Overview Graph, any potential therapy decisions were made based on the incorrect data displayed on the 24-hour sensor glucose overview...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0594-2026
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Medtronic, Inc.
- Product
- CareLink Clinic, REF: MMT-7350
- Product code
- PHV Continuous Glucose Monitor Retrospective Data Analysis Software
- Reason for recall
- Software error causing incorrect data to be displayed on the 24-hour Sensor Glucose Overview Graph, any potential therapy decisions were made based on the incorrect data displayed on the 24-hour sensor glucose overview graph may lead to hypoglycemia or hyperglycemia.
- Root cause tag
- software
- FDA root cause
- Software change control
- Action
- On 10/21/2025, "URGENT MEDICAL DEVICE CORRECTION" notices were emailed to customers informing them of the following: software version 4.2C was released to correct the issues with the data displayed on 24-hour Sensor Glucose Overview graph. If you made individual diabetes treatment recommendations to patients based on the hypoglycemic patterns or on the 30-day sensor glucose data displayed on the 24-hour Sensor Glucose Overview graph with affected software, revisit those recommendations utilizing patients individual CareLink reports, as they were not impacted by this issue. Complete and return the confirmation form: https://info.medtronicdiabetes.com/carelink-clinic-dashboard-hcp or email form to [email protected] For questions, contact the firm's 24-Hour Technical Support line at 1-800-646-4633, option 9.
- Quantity in commerce
- 12126
- Distribution
- Worldwide Distribution: US (nationwide) including states of: TX, NC, MN, CA, VA, ID, WA, GA, MI, MA, OH, IL, IA, CT, FL, PA, NY, TN, AZ, LA, AL, NJ, CO, SC, MS, OR, MT, MD, KS, AK, NH, NE, WY, ND, IN, MO, WI, KY, UT, NV, SD, WV, NM, HI, AR, OK, VT, RI; and OUS (foreign) to countries of: Australia (AU), Sweden (SE), Belgium (BE), Great Britain (GB), Netherlands (NL), Finland (FI), Canada (CA), Germany (DE), Spain (ES), Switzerland (CH), Italy (IT), South Africa (ZA), South Korea (KR), Singapore (SG), New Zealand (NZ), Poland (PL), Israel (IL), Denmark (DK), Norway (NO), Luxembourg (LU), Lithuania (LT), Czech Republic ( CZ), Mexico (MX), Brazil (BR), Colombia (CO), Slovakia (SK), Ireland (IE), Argentina (AR), Greece (GR), Kuwait (KW), Portugal (PT), Iceland (IS), Japan (JP), Turkey (TR), Austria (AT), Qatar (QA), Hong Kong (HK), Chile (CL), Ukraine (UA), Hungary (HU), Estonia (EE), RS, Latvia (LV), Taiwan (TW), Saudia Arabia (SA), Bahrain (BH), Oman (OM), Romania (RO), India (IN), Uruguay (UY), Malaysia (MY), United Arab Emirates (AE), Algeria (DZ), Croatia (HR), France (FR), Russia (RU), Cyprus (CY), Thailand (TH), Egypt (EG), Bosnia and Herzegovina (BA), Libya (LY), and Iraq (IQ)
- Product codes and lots
- UDI-DI: 0763000B00008748K. Software Versions: 4.2B Software Available: US: Between September 4th, 2025, and October 3rd, 2025. OUS: Between September 11th, 2025, and October 4th, 2025. US (Clinical): Between August 21st, 2025, and October 3rd, 2025. OUS (Clinical): Between August 21st, 2025, and October 4th, 2025.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0594-2026 | Class II | Ongoing | Voluntary: Firm initiated |