Z-0531-2026Class I
SJM Pericardial Patch, with EnCap Technology 9x14cm, Model Number REF C0914; PATCH, PLEDGET AND INTRACARDIAC...
Glycar SA Pty., Ltd. / initiated 2025-10-09 / Open, Classified / root cause: manufacturing process
Glycar SA Pty., Ltd. began a recall of SJM Pericardial Patch, with EnCap Technology 9x14cm, Model Number REF C0914; PATCH, PLEDGET AND INTRACARDIAC, PETP, PTFE, POLYPROPYLENE on . FDA classified it as Class I and gives the reason as "The impacted lot may not meet the required tensile strength specification". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0531-2026
- Classification
- Class I
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Glycar SA Pty., Ltd.
- Product
- SJM Pericardial Patch, with EnCap Technology 9x14cm, Model Number REF C0914; PATCH, PLEDGET AND INTRACARDIAC, PETP, PTFE, POLYPROPYLENE
- Product code
- DXZ Patch, Pledget And Intracardiac, Petp, Ptfe, Polypropylene
- Reason for recall
- The impacted lot may not meet the required tensile strength specification.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Glycar issued a Stock Withdrawal Notification to its sole distributor on 10/09/2025. The notice explained the issue and requested the following: "Required Actions 1. Immediately identify and segregate all affected stock in your possession. 2. Clearly label the identified stock as "WITHDRAWN PRODUCT" DO NOT DISTRIBUTE OR USE. 3. Suspend any further distribution or transfer of the affected lots. 4. Provide written confirmation of the quantity and storage location of all affected stock within five (5) business days of receipt of this notification. 5. Our Quality team will coordinate collection, return, or replacement of affected product as appropriate." For questions or require further clarification, contact: Malcolm Applewhite, [email protected] On 10/17/2025, the firm issued an URGENT: MEDICAL DEVICE RECALL notice the medical facility via email. This notice explained the issue, potential risk to health, and requested the following: "Actions to be Taken by the Customer / User: 1. Identify and quarantine any affected products with lot number T2502257 in your possession. 2. Cease use of the affected device immediately. 3. Share this notice with applicable personnel within your institution. 4. Abbott will coordinate and arrange for return and replacement of affected units. 5. Complete and return the attached Acknowledgment Form within 5 business days to confirm receipt of this notice." For questions or assistance, contact: Regulatory Affairs Manager, Glycar SA (Pty) Ltd, Malcolm Applewhite, [email protected]
- Quantity in commerce
- 30 units
- Distribution
- US
- Product codes and lots
- UDI/DI 05414734007276, Lot Numbers: T2502257, exp. 20280128; T2505144, exp. 2028.04.11
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0531-2026 | Class I | Ongoing | Voluntary: Firm initiated | ||
| Z-0532-2026 | Class I | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.