Medline Industries, LP began a recall of Convenience Kits: DYNJ901048N ORTHO SPINE ST CLAIR DYNJ909123F UROLOGY MINOR DYNJ909132F SET UP DYNJ909308D HIP PINNING on . FDA classified it as Class I and gives the reason as "Firm received multiple complaints of anesthesia circuit leaks due to cracks in the 120" expandable tubing. Cracks and resulting leaks may lead to hypoxia, hypoventilation, and inadequate delivery of inhaled anesthetic...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0448-2026
- Classification
- Class I
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- MEDLINE INDUSTRIES, LP - Northfield
- Product
- Convenience Kits: DYNJ901048N ORTHO SPINE ST CLAIR DYNJ909123F UROLOGY MINOR DYNJ909132F SET UP DYNJ909308D HIP PINNING
- Product code
- LRO General Surgery Tray
- Reason for recall
- Firm received multiple complaints of anesthesia circuit leaks due to cracks in the 120" expandable tubing. Cracks and resulting leaks may lead to hypoxia, hypoventilation, and inadequate delivery of inhaled anesthetic agents, which may result in serious injury or death. Volatile anesthetics may also escape into the operating room environment.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Safety notice was provided to customers on 10/10/2025. Follow-up "URGENT MEDICAL DEVICE RECALL" notification dated 10/21/2025 was sent out on 10/21/2025. Notification was through email and first-class mail. Customers were instructed to Immediately check your stock for the affected item number(s) and the affected lot number(s) listed within the recall portal. Quarantine all affected product and destroy affected kits on hand. No products are to be returned to Medline. Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected product in inventory, please complete and submit the response form. Website link: https://recalls.medline.com Recall Reference #: R-25-215-FG Recall Code: If you are a distributor or have resold or transferred this product to another company or individual, you are required by FDA regulations to notify them of this recall communication. Please request that customers destroy any affected product. You should include your customers quantities on your response form. An expanded "URGENT MEDICAL DEVICE RECALL" notification dated 10/24/2025 was provided to customers on 10/24/25, which included additional affected products. If you have any questions, contact the Recall Department at 866-359-1704 or [email protected].
- Quantity in commerce
- 96,406 total units
- Distribution
- International distribution to Mexico and Panama. US nationwide distribution to the following: AK AL AR AZ CA CO CT FL GA IA IL IN KS KY MA ME MI MN MO MS MT NC NJ NY OH OR PA PR RI SC TN TX WA WI;
- Product codes and lots
- HIP PINNING DYNJ909308D UDI-DI 40198459055288 (case), 10198459055287 (ea) Lots 25FBB311 25GBC282 25GBW486; ORTHO SPINE ST CLAIR DYNJ901048N UDI-DI 40198459055554 (case), 10198459055553 (ea) Lots 25GBQ070; SET UP DYNJ909132F UDI-DI 40198459130572 (case), 10198459130571 (ea) Lots 25GBI845 25IBF708; UROLOGY MINOR DYNJ909123F UDI-DI 40198459130602 (case), 10198459130601 (ea) Lots 25GBT826 25IBF707
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0442-2026 | Class I | Ongoing | Voluntary: Firm initiated | ||
| Z-0443-2026 | Class I | Ongoing | Voluntary: Firm initiated | ||
| Z-0444-2026 | Class I | Ongoing | Voluntary: Firm initiated | ||
| Z-0445-2026 | Class I | Ongoing | Voluntary: Firm initiated | ||
| Z-0446-2026 | Class I | Ongoing | Voluntary: Firm initiated | ||
| Z-0447-2026 | Class I | Ongoing | Voluntary: Firm initiated | ||
| Z-0448-2026 | Class I | Ongoing | Voluntary: Firm initiated | ||
| Z-0449-2026 | Class I | Ongoing | Voluntary: Firm initiated | ||
| Z-0450-2026 | Class I | Ongoing | Voluntary: Firm initiated | ||
| Z-0451-2026 | Class I | Ongoing | Voluntary: Firm initiated |