Z-0940-2026Class II
MADSEN ACCUSCREEN TEOAE/DPOAE/ABR Probe (P/N 8-69-41100)
Path / initiated 2025-11-06 / Open, Classified
PATH began a recall of MADSEN ACCUSCREEN TEOAE/DPOAE/ABR Probe (P/N 8-69-41100) on . FDA classified it as Class II and gives the reason as "Following a component change in the DPOAE probe (speaker replacement), technical distortions may occur at high stimulus levels combined with unusually low OAE amplitude acceptance thresholds. These distortions can mimic...". As of , the recall status on record is Open, Classified.
Recall record
- Recall number
- Z-0940-2026
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Path
- Product
- MADSEN ACCUSCREEN TEOAE/DPOAE/ABR Probe (P/N 8-69-41100)
- Product code
- GWJ Stimulator, Auditory, Evoked Response
- Reason for recall
- Following a component change in the DPOAE probe (speaker replacement), technical distortions may occur at high stimulus levels combined with unusually low OAE amplitude acceptance thresholds. These distortions can mimic true DPOAE responses and may result in a false PASS outcome.
- FDA root cause
- Component change control
- Action
- On November 6, 2025 URGENT: MEDICAL DEVICE Field Corrective Action letters were sent to their two distributors. Actions to be taken: - Users will be asked to confirm the DPOAE protocol in use. - Verify the DPOAE Protocol used is not affected by the issue (DP1, DP2, DP3, DP4, DP17, DP18, DP19, DP20, DP21). - If any other protocol is used, please replace it by using AccuLink Software. Follow instructions in IFU to do so. - Remove all affected protocols (DP5-DP16) using AccuLink Software. Follow instructions in IFU to do so. - Complete and return the attached form to confirm these steps were followed. - The expected lifetime of the probes is two years from date of manufacture. Please put the products out of use, once this is reached.
- Distribution
- Domestic: WI. International: Ireland.
- Product codes and lots
- Part Number: 8-69-41100; UDI-DI: 04260223141355; Serial Numbers: 1008852, 1008888, 1008916, 1008925, 1008935, 1008945, 1008951, 1008957, 1008960, 1008962, 1008966, 1008967, 1008987, 1008990, 1009054, 1009065, 1009066, 1009131, 1009137, 1009138, 1009139, 1009145, 1009148, 1009155, 1009159, 1009161, 1009169, 1009184, 1009188, 1009191, 1009305, 1009315, 1009316, 1009319, 1009321, 1009322, 1009325, 1009328, 1009343, 1009352, 1009356, 1009357, 1009359, 1009366, 1009372, 1009375, 1009376, 1009386, 1009388, 1009390, 1009396, 1009399, 1009400, 1009404, 1009411, 1009415, 1009419, 1009426, 1009427, 1009432, 1009439, 1009442, 1009454, 1009456, 1009460, 1009463, 1009464, 1009468, 1009475, 1009481, 1009488, 1009494, 1009495, 1009497, 1009848, 1009855.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0940-2026 | Class II | Ongoing | Voluntary: Firm initiated |