Z-0585-2026Class II
1. Camhb W/lhb, Manual, 11Ml, 10/Box Ycp112-10 2. Brucella Broth, 11Ml, 10/Box Yt3450
Remel Inc / initiated 2025-11-12 / Open, Classified
Remel, Inc began a recall of 1. CAMHB W/LHB, MANUAL, 11ML, 10/BOX YCP112-10 2. BRUCELLA BROTH, 11ML, 10/BOX YT3450 on . FDA classified it as Class II and gives the reason as "Products may contain contamination, which may result in a darker or brown media color". As of , the recall status on record is Open, Classified.
Recall record
- Recall number
- Z-0585-2026
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Remel Inc
- Product
- 1. Camhb W/lhb, Manual, 11Ml, 10/Box Ycp112-10 2. Brucella Broth, 11Ml, 10/Box Yt3450
- Product code
- JTZ Culture Media, Antimicrobial Susceptibility Test, Mueller Hinton Agar/Broth
- Reason for recall
- Products may contain contamination, which may result in a darker or brown media color.
- FDA root cause
- Under Investigation by firm
- Action
- On November 13, 2025, firm began notifying customers via Urgent Medical Device Recall letters. Customers were advised to review results and destroy any remaining inventory of affected lots. Affected product should not be used for any clinical laboratory testing. Customers should complete the acknowledgement form provided by the firm indicating number of units destroyed in order to receive replacement/credit.
- Quantity in commerce
- 1,000 US, 280 ROW
- Distribution
- Worldwide - US Nationwide distribution in the states of AK, CA, FL, GA, IA, IL, IN, LA, MA, MD, ME, MI, MN, NE, NY, OH, OK, PA, RI, SD, TN, VA, WA, WI and the countries of Brazil, Canada, China, Cyprus, Finland, Great Britain, Ireland, Moldova, Norway, Singapore.
- Product codes and lots
- 1. UDI-DI 848838018831, lots 305529, 311228, 316019, 320993 2. UDI-DI 848838091353, lot 317287
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0585-2026 | Class II | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.