Orthofix NorthStar OCT Navigation Surgical Technique Guide, and instruments: Navigation Driver, REF...
SEASPINE ORTHOPEDICS CORPORATION began a recall of Orthofix NorthStar OCT Navigation Surgical Technique Guide, and instruments: Navigation Driver, REF: PC2-400012; Navigation Straight Bone Probe, REF... on . FDA classified it as Class II and gives the reason as "The NorthStar OCT Navigation Surgical Technique Guide (STG) incorrectly identified both Medtronic s Solera and Medtronic s Infinity system toolcards as compatible with the NorthStar OCT Navigation Instruments. However...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-0937-2026
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- SeaSpine Orthopedics Corporation
- Product
- Orthofix NorthStar OCT Navigation Surgical Technique Guide, and instruments: Navigation Driver, REF: PC2-400012; Navigation Straight Bone Probe, REF: PC2-400013; Navigation, 3.5mm Drill, REF: PC2-400135; Navigation, 4.0mm Drill, REF: PC2-400140; Navigation, 4.5mm Drill, PC2-400145; Navigation, 5.5mm Drill, REF: PC2-400155; Navigation, 3.5mm Tap, REF: PC2-400235; Navigation, 4.0mm Tap, REF: PC2-400240; Navigation, 4.5mm Tap, REF: PC2-400245; Navigation, 4.5mm Dual Lead Tap, REF: PC2-400345; Navigation, 5.5mm Dual Lead Tap, REF: PC2-400355
- Product code
- OLO Orthopedic Stereotaxic Instrument
- Reason for recall
- The NorthStar OCT Navigation Surgical Technique Guide (STG) incorrectly identified both Medtronic s Solera and Medtronic s Infinity system toolcards as compatible with the NorthStar OCT Navigation Instruments. However, the NorthStar OCT Navigation Instruments are compatible for use with Medtronic s Solera toolcards only.
- Root cause tag
- labeling and UDI
- FDA root cause
- Labeling design
- Action
- On 10/27/2025, device correction notices were emailed to customers who were informed of the following: The NorthStar OCT Navigation Surgical Technique Guide (STG) incorrectly identified both Medtronic's Solera and Medtronics Infinity system StealthStation tool cards as compatible with the NorthStar OCT Navigation Instruments; however, the NorthStar OCT Navigation Instruments are indicated for use with Medtronic's Solera tool cards only. 1) Discontinue use of Medtronic's Infinity tool cards with NorthStar OCT Navigation instruments. 2) Discard affected version of the Surgical Technique Guide 3) This notice should be shared within your organization and forwarded to any organization where affected Instruments and guide is utilized. 4) Review updated Surgical Technique Guide D0003926 Revision DA. This notice should be shared within your organization and forwarded to any organization where affected instruments are utilized. Any product complaints should be reported to Orthofix at [email protected]
- Quantity in commerce
- 310
- Distribution
- US: CT, TX, NV, CO, CA, WA, NJ, OR, IN, MI, FL, PA, IL, VA, AL, RI, MO, AZ, MS, SC, UT, NY, OK, SD. OUS: Australia, Mexico, Chile
- Product codes and lots
- Surgical Technique Guide revisions: D0003926A, D0003926B, D0003926C, D0003926D. REF/UDI-DI/Lot: PC2-400012/10889981260033/MK81527A, MK81527ARR, MK81527ARRR, MK81574AR, MK81574AR1, MK93487A, MK93488ARR, MK93489A, MK93490A, MK93490ARR, MK93491AR, MK93491AR1, MK943488A, MK95882A, MK95882ARR, ML52083B, ML53136B, ML53188B, ML69777B, ML69778B, ML71257B, MM1001498D, MM1001499D, MM1001748D, MM1001749D, MM1001813D, MM1001902D, MM1001903D, MM1002025D, MM1002316D, MM1002328D, MM1002344D; PC2-400013/10889981260040/AL657979C, AL658549C, AL658549CRRR, AL658641C, AL661470C, AL671917D, AL683456F, AL702785F; PC2-400135/10889981260057/BK104899A, BK104899AR, BK104899ARR, BK104899ARRR, BK104899ARRR4, BK106822A, CR49020D, CR49023D, CR49060D, CR51550E, CR51972D, CR56472E; PC2-400140/10889981260064/BK104900A, BK104900AR, BK104900ARRR, BK106825A, BK107806A, BK107806ARRR, BK108366A, BK108366ARRR, CR49822D, CR50158D, CR51973D, CR57547E, CR59411E; PC2-400145/10889981260071/BK104901A, BK104901ARRR, BK106827A, CR49021C, CR49061C, CR51974C; PC2-400155/10889981260088/BK104902A, BK104902ARRR, BK106829A, CR49062C, CR51551C, CR51975C, CR57550E, CR59413E; PC2-400235/10889981260095/BK104903A, BK104903ARRR, BK106831A, BK106831ARRR, BK131159C, CR49620C, CR52371D, CR54729D, CR58013D, CR59414D; PC2-400240/10889981260101/BK104904A, BK104904ARRR, BK106833A, BK131160C, CR49621C, CR52372D, CR59415D; PC2-400245/10889981275495/no lots released; PC2-400345/10889981260118/BK104905A, BK104905ARRR, BK106834A, BK106834ARRR, BK131161C, CR49622C, CR52373D, CR54208D; PC2-400355/10889981260125/BK104906A, BK106835A, CR51764C, CR52853D
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0937-2026 | Class II | Ongoing | Voluntary: Firm initiated |