Brand Name: Single Use Electrosurgical Knife KD-645 Product Name: Olympus Single Use Electrosurgical Knife...
Olympus Corporation of the Americas began a recall of Brand Name: Single Use Electrosurgical Knife KD-645 Product Name: Olympus Single Use Electrosurgical Knife KD-645 Model/Catalog Number: KD-645L Product... on . FDA classified it as Class II and gives the reason as "A deterioration of the cutting knife, including overheating and burning, can contribute to tip breakage during use". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1223-2026
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- OLYMPUS Corporation of the Americas
- Product
- Brand Name: Single Use Electrosurgical Knife KD-645 Product Name: Olympus Single Use Electrosurgical Knife KD-645 Model/Catalog Number: KD-645L Product Description: An electrosurgical device consisting of a handpiece with mechanical and/or electrical controls, and a monopolar electrode, intended to deliver a current from a system generator directly to tissues for cutting/coagulation/ablation during endoscopic (including laparoscopic, arthroscopic) surgery.
- Product code
- KNS Unit, Electrosurgical, Endoscopic (With Or Without Accessories)
- Reason for recall
- A deterioration of the cutting knife, including overheating and burning, can contribute to tip breakage during use.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- On December 17, 2025 URGENT: MEDICAL DEVICE CORRECTION letters were sent to customers. Actions to be taken: 1. Carefully read the content of this notification. 2. Ensure all personnel are completely knowledgeable and thoroughly trained on the content of this notification, including the instruction for intermittent output activation and applicable Warnings and Cautions. 3. Keep a copy of this notification with the Instructions for Use for any affected devices remaining in your inventory. 4. If you have further distributed the affected product, please identify your customers and forward this notification to them. Olympus fully appreciates your prompt cooperation in addressing this situation. If you require additional information, please do not hesitate to contact me directly by phone at (647) 999-3203 or by e-mail [email protected].
- Quantity in commerce
- 3361 units
- Distribution
- US Nationwide distribution.
- Product codes and lots
- Model/Catalog Number: KD-645L UDI: 04953170407857 Lot Number(s): 2ZK, 31K, 32K, 33K, 35K, 3XK, 3YK, 3ZK, 42K, 43K, 44K, 45K, 46K, 47K, 48K, 49K, 4XK, 4YK, 4ZK, 51K, 52K, 53K, 54K, 55K, 56K, 57K, 58K, 59K
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1222-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1223-2026 | Class II | Ongoing | Voluntary: Firm initiated |