Brand Name: Affixus Antegrade Femoral Nailing System Product Name: Affixus Antegrade Femoral Nails...
Zimmer, Inc. began a recall of Brand Name: Affixus Antegrade Femoral Nailing System Product Name: Affixus Antegrade Femoral Nails Model/Catalog Number: 815509440 Software Version: N/A... on . FDA classified it as Class II and gives the reason as "Affected implants may have undersized distal diameter along length of the shaft, which could lead to implant fatigue fracture. Potential health risk if implant fatigue fracture occurs include adverse local tissue...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0973-2026
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Zimmer GmbH
- Product
- Brand Name: Affixus Antegrade Femoral Nailing System Product Name: Affixus Antegrade Femoral Nails Model/Catalog Number: 815509440 Software Version: N/A Product Description: Affixus 9 mm, Length 440 mm, Right, Trochanteric Nail Component: N/A
- Product code
- HSB Rod, Fixation, Intramedullary And Accessories
- Reason for recall
- Affected implants may have undersized distal diameter along length of the shaft, which could lead to implant fatigue fracture. Potential health risk if implant fatigue fracture occurs include adverse local tissue reaction, pain or ache, non-union or malunion, or tissue damage necessitating surgical intervention.
- Root cause tag
- manufacturing process
- FDA root cause
- Process design
- Action
- Firm began notifying customers on December 2, 2025 via Urgent Medical Device Recall letters. Distributors are to locate and quarantine affected product in their inventory. They must also return all affected product from their distributorship and from affected hospitals within their territory. Risk Managers are to assist Zimmer Biomet sales representatives in quarantining all affected product. Sales representatives will remove the product from customer facilities. Surgeons are to maintain awareness of the issue, but there are no patient-monitoring instructions related to this recall recommended beyond their existing follow-up schedules.
- Quantity in commerce
- 22
- Distribution
- US Nationwide distribution in the states of AL, AZ, CA, FL, IN, KY, MA, MO, MT, NC, NM, RI, TX, UT.
- Product codes and lots
- Lot Code: Model No 815509440 ; UDI-DI (01)00887868143588(17)340820(10)66717557 ; Lot Number 66717557 Model No 815509440 ; UDI-DI (01)00887868143588(17)350422(10)67101279 ; Lot Number 67101279 Model No 815509440 ; UDI-DI (01)00887868143588(17)350419(10)67101280 ; Lot Number 67101280
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0967-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0968-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0969-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0970-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0971-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0972-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0973-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0974-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0975-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0976-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0977-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0978-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0979-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0980-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0981-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0982-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0983-2026 | Class II | Ongoing | Voluntary: Firm initiated |