Brand Name: Molift Product Name: Molift 2-point sling bars (Accessory to mobile hoists) Model/Catalog...
ETAC A/S began a recall of Brand Name: Molift Product Name: Molift 2-point sling bars (Accessory to mobile hoists) Model/Catalog Numbers: (1) 1530101, (2) 1530102 Product Description... on . FDA classified it as Class II and gives the reason as "A customer experienced that during assembly and load testing of the device, the bolt for mounting the lifting bar broke. Upon further investigation of this customer complaint, it became clear that the defect occurs as a...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1310-2026
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Etac A/S
- Product
- Brand Name: Molift Product Name: Molift 2-point sling bars (Accessory to mobile hoists) Model/Catalog Numbers: (1) 1530101, (2) 1530102 Product Description: The mobile hoist is an assistive device intended for alleviation of, or compensation for, a functional, impairment due to an injury or disability. The device is designed for an individual lacking the ability to stand up and transfer themselves to / from a bed, the floor, a wheelchair, a chair, a toilet or similar due to reduced mobility or physical strength.
- Product code
- FSA Lift, Patient, Non-Ac-Powered
- Reason for recall
- A customer experienced that during assembly and load testing of the device, the bolt for mounting the lifting bar broke. Upon further investigation of this customer complaint, it became clear that the defect occurs as a result of deformation of the materials involved between the lifting bar and the arm of the device. Such deformation can lead to limited mobility in the joint, which prevents proper functionality of the equipment.
- Root cause tag
- design
- FDA root cause
- Component design/selection
- Action
- On November 17, 2025 customers were notified vis a phone call. Notifying them that they will receive an email with the Field Safety Notice and Response Form. The customers were emailed Field Safety Notice letters on or around November 20, 2025. Corrective action: To reduce the risk of further incidents, replacement units are offered for all potentially affected products. Continued use is safe as long as the lifting bar can move freely and remains vertical during the lifting cycle with or without load. If this is not the case, or if there is any doubt about the product s functionality, the unit must be taken out of service immediately. The components in the defective units (Mover 205 and Mover 300) will gradually be replaced in the coming months. The affected lifting bars will likewise be replaced.
- Quantity in commerce
- 64 units
- Distribution
- Worldwide - US Nationwide distribution in the state of PA and the countries of Australia, Czech Republic, Finland, France, Germany, Israel, Italy, Japan, New Zealand, Norway, Poland, Qatar Spain, Sweden, Switzerland, UK.
- Product codes and lots
- (1) Model/Catalog Number: 1530101; UDI-DI: TBD; (2) Model/Catalog Number: 1530102; UDI-DI: TBD;
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1308-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1309-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1310-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1311-2026 | Class II | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.
