Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1037-2026Class II

Codman CereLink ICP Sensor Basic Kit. Catalog Number: 826850. The Codman CereLink ICP Sensor Basic Kit...

Integra LifeSciences Corporation / initiated 2025-12-12 / Open, Classified / root cause: manufacturing process

Integra LifeSciences Corp. (NeuroSciences) began a recall of Codman CereLink ICP Sensor Basic Kit. Catalog Number: 826850. The Codman CereLink ICP Sensor Basic Kit consists of the CereLink ICP Sensor (ICP Sensor) and a... on . FDA classified it as Class II and gives the reason as "Potential corrosion stains on the surface of the 14-gauge Tuohy Needle that is included in the Microsensor and Cerelink ICP Kits". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1037-2026
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Integra LifeSciences Corporation
Product
Codman CereLink ICP Sensor Basic Kit. Catalog Number: 826850. The Codman CereLink ICP Sensor Basic Kit consists of the CereLink ICP Sensor (ICP Sensor) and a 14-gauge Tuohy needle with stylet. The ICP Sensor is a nylon tube with a microminiature strain gauge pressure transducer (sensing element) mounted at one end and an electrical connector at the other end. It is designed for use with a Codman intracranial pressure monitoring device. The ICP Sensor Basic Kit is designed for use with the Codman Cranial Hand Drill. The drill facilitates access to the intraparenchymal area. The drill is also available as a component of the Codman Cranial Access Kit. Use of the ICP Sensor Basic Kit is indicated when direct ICP monitoring is required. The kit is indicated for use in both subdural and intraparenchymal pressure monitoring applications only.
Product code
GWM Device, Monitoring, Intracranial Pressure
Reason for recall
Potential corrosion stains on the surface of the 14-gauge Tuohy Needle that is included in the Microsensor and Cerelink ICP Kits.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Integra LifeSciences Corporation notified consignees on about 12/12/2025 via FedEx overnight mailing. Consignees were instructed to identify impacted part numbers and lot numbers on hand, remove them immediately from service and quarantine them, complete and return the Acknowledgement Form, and maintain the notice for records. Consignees were also instructed to forward the notification to those who utilize the product for awareness of the recall. Once the Acknowledgement Form is received, and noted that affected product are on hand, Customer Service will contact consignees and provide a Return Material Authorization (RMA) number and directions to return the affected product.
Quantity in commerce
5,116 units
Distribution
Worldwide - US Nationwide distribution in the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX UT, VA WA and the countries of Argentina, Australia, Austria, Bangladesh, Belgium, Bolivia (Plurinational State of), Brazil, Canada, Chile, China, Colombia, Cyprus, Czechia, France, Georgia, Germany, Greece, Hong Kong, Iceland, India, Ireland, Israel, Italy, Japan, Kenya, Korea (the Republic of), Lithuania, Nepal, New Zealand, Poland, Portugal, Romania, Saudi Arabia, Singapore, Slovenia, South Africa, Spain, Sweden, Taiwan (Province of China), Thailand, Turkey, United Arab Emirates (the), United Kingdom of Great Britain and Northern Ireland (the).
Product codes and lots
Catalog Number: 826850. UDI-DI: 10381780520672. Lot Numbers: 7261958, 7266691, 7266692, 7288060, 7293837, 7293836, 7306040, 7306041, 7306042, 7306043, 7306044, 7306045, 7306046, 7306047, 7306050, 7306051, 7306052, 7307670, 7307671, 7307672, 7325482, 7307675, 7325481, 7325483, 7325484, 7307673, 7307674, 7325485, 7325486, 7336611, 7336612, 7336614, 7336623, 7336613, 7336615, 7336616, 7336618, 7336619, 7336620, 7336621, 7336622, 7306048, 7306049, 7336626, 7336627, 7336629, 7347166, 7336628, 7346295, 7346297, 7346299, 7346301, 7365434, 7365435, 7392019, 7396607, 7396608, 7396609, 7396610, 7392034, 7392067, 7405790, 7405791, 7405792, 7405793, 7406633, 7406634, 7406636, 7406637, 7406639, 7406640, 7406641, 7406642, 7406635, 7406638, 7408813, 7408815, 7408818, 7408814, 7408816, 7468726, 7468727, 7411801, 7411802, 7414641, 7414642, 7468731, 7468732, 7468745, 7468746, 7468741, 7468742, 7468743, 7468744, 7468771, 7468769, 7468770, 7468773, 7468774, 7468775, 7482179, 7482180, 7482181, 7482182, 7482183, 7482184, 7482185, 7482186, 7468772, 7482154, 7482155, 7482156, 7482157, 7468801, 7468804, 7468805, 7468806, 7482136, 7468802, 7468803, 7482132, 7482134, 7482135, 7482137, 7482159, 7482158, 7482160, 7482161, 7495593, 7495594, 7495596, 7495597, 7495598, 7495599, 7482212, 7495600, 7495601, 7495602, 7495603, 7495604, 7495605, 7495606, 7495607, 7495608, 7495635, 7495636, 7495637, 7495638, 7495639, 7495640, 7495641, 7495642, 7495643, 7495666, 7495644, 7495664, 7482213, 7495667, 7503597, 7503598, 7503599, 7503659, 7510140, 7510141, 7510142, 7503600, 7503601, 7517716, 7517708, 7503627, 7503628, 7503630, 7503631, 7503632, 7503633, 7503629, 7503664, 7503657, 7503662, 7503663, 7503602, 7503624, 7510156, 7517703, 7517704, 7517706, 7517707, 7517709, 7517710, 7517711, 7517712, 7517714, 7517715, 7408817, 7411797, 7411798, 7411799, 7411800, 7414643, 7414644, 7414645, 7414646, 7414647, 7414648, 7414649, 7417938, 7417940, 7417942, 7417944, 7417946, 7417948, 7417950, 7417952, 7417955, 7417957, 7417959, 7503625.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1037-2026
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1036-2026Class IIOngoingVoluntary: Firm initiated
Z-1037-2026Class IIOngoingVoluntary: Firm initiated