A71200 Vanta" Clinician Programmer Application (CP App) version 2.0.2465 and 2.0.2683, Product number A71200...
Medtronic Neuromodulation began a recall of A71200 Vanta" Clinician Programmer Application (CP App) version 2.0.2465 and 2.0.2683, Product number A71200. Software is deployed and operated on Samsung... on . FDA classified it as Class II and gives the reason as "Complaints received that Vanta A71200 CP App does not function as intended during use. There is a potential for prolonged or postponed surgical procedure". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1043-2026
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Medtronic Neuromodulation
- Product
- A71200 Vanta" Clinician Programmer Application (CP App) version 2.0.2465 and 2.0.2683, Product number A71200. Software is deployed and operated on Samsung tablets (CFN CT900).
- Product code
- LGW Stimulator, Spinal-Cord, Totally Implanted For Pain Relief
- Reason for recall
- Complaints received that Vanta A71200 CP App does not function as intended during use. There is a potential for prolonged or postponed surgical procedure.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- On December 3, 2025 the firm distributed Urgent Medical Device Safety Notification letters to affected consignees. Customers were instructed to download and install A81200 CP App v2.0.2684 or any later version.
- Quantity in commerce
- 7123
- Distribution
- Worldwide - US Nationwide distribution.
- Product codes and lots
- version 2.0.2465 and 2.0.2683 UDI-DI (GTIN) 00763000520083 Affected product is the software. Implanted Vanta Neurostimulator devices are not affected.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1043-2026 | Class II | Ongoing | Voluntary: Firm initiated |