Z-1040-2026Class II
Detachable 10in Pend FRU-1 Pend component for the AIRO Computed Tomography (CT) X-ray System
Brainlab, Inc. / initiated 2025-12-04 / Open, Classified / root cause: labeling and UDI
Mobius Imaging, LLC began a recall of Detachable 10in Pend FRU-1 Pend component for the AIRO Computed Tomography (CT) X-ray System on . FDA classified it as Class II and gives the reason as "AIRO pendant replacement kits and upgrade do not possess the "Safe Hand Position Warning" label". As of , the recall status on record is Open, Classified, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-1040-2026
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Brainlab, Inc.
- Product
- Detachable 10in Pend FRU-1 Pend component for the AIRO Computed Tomography (CT) X-ray System.
- Product code
- JAK System, X-Ray, Tomography, Computed
- Reason for recall
- AIRO pendant replacement kits and upgrade do not possess the "Safe Hand Position Warning" label.
- Root cause tag
- labeling and UDI
- FDA root cause
- Labeling Change Control
- Action
- An URGENT MEDICAL DEVICE CORRECTION notice dated 12/4/25 instructs consignees to ensure the following mitigations are in place while the correction is deployed: the hand position warnings as indicated in the IFU, train operators, the AIRO is accessible through a physical key switch and password protected log in screen, and the Backing Up System Warning label is affixed on to the gantry of affected devices. Stryker will correct devices during an upcoming service visit. Consignees are to inform users of devices of this recall by sharing the recall notice, immediately check stock areas and operating rooms to identify affected devices, and inform all users of the potential harms and their risk mitigations. Consignees with any questions are to contact their local sales office or Stryker Representative.
- Quantity in commerce
- 101 units
- Distribution
- US Nationwide distribution in the states of AK, AL, AZ, CA, CO, CT, DC, FL, IL, KY, MA, MD, MI, MN, MO, NC, NY, OH, PA, RI, SC, TN, TX, UT, VA, WI.
- Product codes and lots
- Part No. MI-76-0237; UDI: (01) 00869346000200 (11)230920 (21) AIRO-XXXX; Serial No. 2232206713, 2232206723, 2232206753, 2232206763, 2303430363, 2303430373, 2303430383, 2303430393, 2304034583, 2304034593, 2304034603, 2304034613, 2304034623, 2305514123, 2305514133, 2305514143, 2305514153, 2311019263, 2315101273, 2315101283, 2315101293, 2315101303, 2316030903, 2316030913, 2316703793, 2316703803, 2316703813, 2316703823, 2317415513, 2317415523, 2317415533, 2317415543, 2317415553, 2317905083, 2317905093, 2317905103, 2317905113, 2317905123, 2323513673, 2323513683, 2323513693, 2331419663, 2331419673, 2332117403, 2332117413, 2332117423, 2332117433, 2332117443, 2333404593, 2333404603, 2335506763, 2335506773, 2401909563, 2401909573, 2401909583, 2401909593, 2401909603, 2401909613, 2402911603, 2402911613, 2402911623, 2402911633, 2404302243, 2404302253, 2404302263, 2404302273, 2409620273, 2409620283, 2409620293, 2409620303, 2422207853, 2422207863, 2422207873, 2422207883, 2422207893, 2422207903, 2422207913, 2426303233, 2426303243, 2426303253, 2426303263, 2426303273, 2426303283, 2431210783, 2431210793, 2431210803, 2431210813, 2431210823, 2434504553, 2434504563, 2434504573, 2502103553, 2502103583, 2502103593, 2503121733, 2503121743, 2503121753, 2503616113, 2503616123, 2503616133, 2503616213.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1040-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1041-2026 | Class II | Ongoing | Voluntary: Firm initiated |