Z-1101-2026Class II
Portex Spinal Tray, Item No. A3729-24 A3595
Abbott Medical / initiated 2025-12-19 / Open, Classified
ICU Medical Inc. began a recall of Portex Spinal Tray, Item No. A3729-24 A3595 on . FDA classified it as Class II and gives the reason as "Affected lots of syringes may have a crack on the syringe collar, which may lead to delayed or repeated blood draw/fluid delivery. This issue may also lead to blood leakage during sampling of arterial blood or leakage...". As of , the recall status on record is Open, Classified.
Recall record
- Recall number
- Z-1101-2026
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Abbott Medical
- Product
- Portex Spinal Tray, Item No. A3729-24 A3595
- Product code
- CAZ Anesthesia Conduction Kit
- Reason for recall
- Affected lots of syringes may have a crack on the syringe collar, which may lead to delayed or repeated blood draw/fluid delivery. This issue may also lead to blood leakage during sampling of arterial blood or leakage during fluid delivery.
- FDA root cause
- Under Investigation by firm
- Action
- On December 19. 2025, the firm issued Urgent Medical Device Correction letters to affected customers. Customers were informed that affected lots may have a crack present on the syringe collar. Customers were instructed to discontinue use of affected lots, and to destroy products following their institution's process for destruction. If destruction is not immediately possible, affected product should be quarantined until disposal.
- Quantity in commerce
- 64290 total
- Distribution
- US Nationwide distribution in the states of CA, FL, IL, IN, MA, MI, MN, MO, NE, NJ, NY, OH, OK, TX, WA.
- Product codes and lots
- 1. A3729-24 UDI-DI 30351688075691 Lots 6146204 (corrected on 1/22/26, removed "-LSM" suffix. 2. A3595 UDI-DI 35019517108211 Lot 6157980
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1100-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1101-2026 | Class II | Ongoing | Voluntary: Firm initiated |