Z-1128-2026Class II

Merit Medical , Custom Procedure Kit Reference Numbers: K12T-04270A K12T-07956 K12T-10922B

Merit Medical Systems, Inc. / initiated 2025-12-04 / Open, Classified / root cause: manufacturing process

Merit Medical Systems, Inc. began a recall of Merit Medical , Custom Procedure Kit Reference Numbers: K12T-04270A K12T-07956 K12T-10922B on . FDA classified it as Class II and gives the reason as "Inflation device handle may detach from the syringe during procedure". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1128-2026
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Merit Medical Systems, Inc.
Product
Merit Medical , Custom Procedure Kit Reference Numbers: K12T-04270A K12T-07956 K12T-10922B
Product code
OEQ Angiography/Angioplasty Kit
Reason for recall
Inflation device handle may detach from the syringe during procedure.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
On December 4, 2025 Merit Medical issued a "Urgent Medical Device Recall Notice" to affected consignees via E-Mail. Merit asked consignees to take the following actions: 1. Please read and understand the recall notification. 2. Should the handle detach during use and the balloon not completely deflate, pull the handle out of the inflation syringe, disconnect the inflation syringe from the catheter, and use an alternate syringe to generate vacuum and aspirate residual contrast 3. Ensure that applicable personnel within your organization are made aware of this field action. 3. If the product has been further distributed to other facilities, institutions, or manufacturers, please ensure this notice is immediately shared with them. 4. Please fill out, scan, and email the completed Customer Response Form to Customer Service at [email protected] within seven (7) calendar days
Distribution
Worldwide - Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, VI, VT, WA, WI, WV, WY, St. Thomas , Puerto Rico and the countries of Albania, Algeria, Angola, Argentina, Australia, Austria, Bahrain, Barbados, Belgium, Brazil, Bulgaria, Cambodia, Canada, Cayman Islands, Chile, China, Colombia, Cote d'Ivoire, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Finland, France, Germany, Greece, Guatemala, Honduras, Hong Kong, Iceland, India, Indonesia, Ireland, Israel, Italy, Jamaica, Japan, Kazakhstan, Korea, Republic of Kuwait, Kyrgyzstan, Latvia, Lebanon, Libya, Luxembourg, Madagascar, Mexico, Monaco, Nepal, Netherlands, New Zealand, Nicaragua, Nigeria, Norway, Oman, Pakistan, Panama, Paraguay, Peru, Poland, Portugal, Qatar, Romania, Russian Federation, Rwanda, San Marino, Saudi Arabia, Serbia, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Trinidad and Tobago, Uganda, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Venezuela, Bolivarian, Republic of Viet Nam.
Product codes and lots
Reference Number/UDI Number/Lot Number(s) K12T-04270A: UDI: 00884450389683/ Lot: T2643938 T2728112 T2783917 T2806210 T2811252 K12T-07956: UDI: 00884450264003/ Lot: K2618155 K2618156 K12T-10922B: UDI: 00884450481660/Lot: T2851185 T2876341 T2990792 T3063197 T3070788 T3165727 T3223759 T3312055

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1128-2026
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1123-2026Class IIOngoingVoluntary: Firm initiated
Z-1124-2026Class IIOngoingVoluntary: Firm initiated
Z-1125-2026Class IIOngoingVoluntary: Firm initiated
Z-1126-2026Class IIOngoingVoluntary: Firm initiated
Z-1127-2026Class IIOngoingVoluntary: Firm initiated
Z-1128-2026Class IIOngoingVoluntary: Firm initiated
Z-1129-2026Class IIOngoingVoluntary: Firm initiated