Z-1130-2026Class II
Brand Name: LEASEIR Product Name: LEASEIR MHR XCELL Model/Catalog Number: MHR 110-b Software Version: 1.5.0.0...
Leaseir Technologies SLU / initiated 2025-12-15 / Completed / root cause: labeling and UDI
LEASEIR TECHNOLOGIES SLU began a recall of Brand Name: LEASEIR Product Name: LEASEIR MHR XCELL Model/Catalog Number: MHR 110-b Software Version: 1.5.0.0 Product Description: The Leaseir MHR Xcell is a... on . FDA classified it as Class II and gives the reason as "The console label for affected devices is missing the "DANGER" symbol". As of , the recall status on record is Completed, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-1130-2026
- Classification
- Class II
- Status
- Completed
- Initiated
- Posted
- Recalling firm
- Leaseir Technologies SLU
- Product
- Brand Name: LEASEIR Product Name: LEASEIR MHR XCELL Model/Catalog Number: MHR 110-b Software Version: 1.5.0.0 Product Description: The Leaseir MHR Xcell is a surgical laser instrument for use in general and plastic surgery and dermatology, intended for hair removal and treatment of pseudo folliculitis barbae on all skin types (Fitzpatrick I-VI). The process implies the generation of intense light pulses at specific wavelengths. The specific nature of the energies and pulse durations cause the desired effect (in the hair follicles and oxyhemoglobin in the blood) heating them sufficiently and destroying them without unnecessary damage to surrounding tissue. This is achieved by controlled emission of laser radiation to the target tissue. The system consists of a main console and two interchangeable applicators that delivers pulsed light in the range of 800-820nm with a peak in the 810nm in different spot sizes. Two of the applicators emitting radiation at 810 nm wavelength for two areas 13.515 mm and 14.525.5 mm. Two different operation modes are available: static mode and dynamic mode, which are differentiated basically by the frequency range defined for each mode (1-4 Hz for static and 10 Hz for dynamic). The principle of operation consists in the photons travel along the axis and reflected again back into the crystal, continuing the chain reaction, while photons travelling in different directions leave the crystal. In one of the two mirrors, a tiny hole allows a small amount of light to leak out and the resulting beam is focused with lenses and is emitted from the laser. The total energy emitted by the Leaseir MHR Xcell is produced by an array of diodes. The laser emission is activated by the hand piece trigger, deliver a continuous pulse pattern while the button is pressed. Component: NO
- Product code
- GEX Powered Laser Surgical Instrument
- Reason for recall
- The console label for affected devices is missing the "DANGER" symbol.
- Root cause tag
- labeling and UDI
- FDA root cause
- Error in labeling
- Action
- On October 29, 2025, the firm notified the single US consignee. This was followed up by an email on December 18, 2025. The consignee was given information regarding the field corrective action for the console labels and missing danger symbol. The consignee was provided with replacement labels and instructions for the correction of both affected units.
- Quantity in commerce
- 2
- Distribution
- US Nationwide distribution in the states of Austin, Texas.
- Product codes and lots
- Lot Code: Device#1: Model: Leaseir MHR Xcell Serial Number: C10016, C10017 UDI-DI: 08437019546076
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1130-2026 | Class II | Completed | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY