Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1225-2026Class II

PRISMAFLEX ST150 SET US, Product Code 107640US; Dialyzer, High Permeability With Or Without Sealed Dialysate...

Baxter Healthcare Corporation / initiated 2026-01-06 / Open, Classified / root cause: design

VANTIVE US HEALTHCARE LLC began a recall of PRISMAFLEX ST150 SET US, Product Code 107640US; Dialyzer, High Permeability With Or Without Sealed Dialysate System on . FDA classified it as Class II and gives the reason as "There is a potential dislodgement of the Prismaflex Set deaeration chamber from the Prismaflex Control Unit". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.

Recall record

Recall number
Z-1225-2026
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Baxter Healthcare Corporation
Product
PRISMAFLEX ST150 SET US, Product Code 107640US; Dialyzer, High Permeability With Or Without Sealed Dialysate System
Product code
KDI Dialyzer, High Permeability With Or Without Sealed Dialysate System
Reason for recall
There is a potential dislodgement of the Prismaflex Set deaeration chamber from the Prismaflex Control Unit.
Root cause tag
design
FDA root cause
Component design/selection
Action
Vantive issued an Urgent Medical Device Correction notice to its consignees on 01/06/2026 via USPS first class mail. The notice explained the issue, hazard involved, and requested the following: 1. You may continue to perform therapy using the Prismaflex sets listed above. Please monitor the deaeration chamber during therapy to ensure it remains in an upright position within the holder. 2. Vantive is aware of customers who have experienced this issue and attempted to further secure the deaeration chamber in an upright position. If attempting to secure the deaeration chamber in an upright position, please consider the following: . Ensure there are no kinks in the tubing and that the deaeration chamber remains visible. . If the Prismaflex Control Unit issues an Air in Blood alarm, check if air is present. If there is no air present, check if the deaeration chamber is dislodged from the holder and ensure there are no clots present. 1. If no dislodgement, please follow the operator s manual instructions for Air in Blood alarm. 2. In case of dislodgement with clotting, follow the associated instructions within the Prismaflex Control Unit operator s manual to stop the therapy and replace the set. 3. Only in case of dislodgement without clotting or any other alarms, if the chamber is secured in an upright position, follow the associated instructions within the Prismaflex Control Unit operator s manual to proceed with treatment. Facilities who further distributed the products were directed to notify those departments or facilities. Dealers, wholesalers, distributor/resellers, or original equipment manufacturers (OEM) who further distributed the products were directed to notify their customers. For general questions regarding this communication, please contact Vantive Customer Care at 833-542-CRRT (2778). Vantive Customer Care is available 24 hours a day, seven days a week.
Quantity in commerce
264,040 units
Distribution
US Nationwide distribution.
Product codes and lots
UDI/DI 00085412917696, All lots including and manufactured after 24F0085

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1225-2026
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1224-2026Class IIOngoingVoluntary: Firm initiated
Z-1225-2026Class IIOngoingVoluntary: Firm initiated
Z-1226-2026Class IIOngoingVoluntary: Firm initiated
Z-1227-2026Class IIOngoingVoluntary: Firm initiated
Z-1228-2026Class IIOngoingVoluntary: Firm initiated
Z-1229-2026Class IIOngoingVoluntary: Firm initiated
Z-1230-2026Class IIOngoingVoluntary: Firm initiated
Z-1231-2026Class IIOngoingVoluntary: Firm initiated