Z-1116-2026Class I
Brand Name: Broselow Product Name: Broselow Domestic Complete ALS Organizer Model/Catalog Number: 7730ALS...
Salter Labs / initiated 2025-12-15 / Open, Classified / root cause: design
SunMed Holdings, LLC began a recall of Brand Name: Broselow Product Name: Broselow Domestic Complete ALS Organizer Model/Catalog Number: 7730ALS Software Version: Not Applicable Product Description... on . FDA classified it as Class I and gives the reason as "Product contains potentially harmful dosing errors for Vecuronium, Flumazenil, and Ketamine". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1116-2026
- Classification
- Class I
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Salter Labs
- Product
- Brand Name: Broselow Product Name: Broselow Domestic Complete ALS Organizer Model/Catalog Number: 7730ALS Software Version: Not Applicable Product Description: The Broselow Tape is a length- and weight-based reference tool specifically designed for pediatric emergency care. Component: No
- Product code
- OKI Emergency Response Safety Kit
- Reason for recall
- Product contains potentially harmful dosing errors for Vecuronium, Flumazenil, and Ketamine.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- AirLife issued an URGENT: MEDICAL DEVICE RECALL NOTICE to its consignees on 12/15/2025 via email and US mail. The notice explained the issue, potential risk to the patient, and requested the following: "Customer immediate actions: Please take the following actions immediately: 1. Stop using the affected tapes o Immediately discontinue use of all AirLife Broselow Rainbow Tapes identified as 2025 Edition, 36-23446 Rev 3 Print Version. 2. Identify and segregate affected product o Examine your inventory and clinical areas for affected Broselow Rainbow Tapes (see Image B for identifying characteristics). o Remove all affected tapes from clinical service. o Segregate or quarantine the affected tapes to prevent further use. 3. Follow disposition instructions o Follow the instructions provided in the accompanying communication regarding discarding/destruction of the affected tapes. o Do not redistribute or place any affected tapes back into service. 4. Notify your internal users o Ensure that all clinicians and healthcare professionals within your organization who may use the Broselow Rainbow Tape are informed of this recall/field removal and understand that the affected Rev 3 tapes must not be used. 5. Notify downstream customers (if applicable) o If you have further distributed the affected tapes, please identify your customers/consignees and notify them of this recall/field removal promptly. o Your notification may be enhanced by including a copy of this letter. o If you have impacted product on hand, do not ship it; instead, hold it for discarding/destruction as instructed. 6. Complete and return the Response Form o Please complete and return the attached Response Form via e-mail to [email protected] as soon as possible. Please complete and return Attachment B if you have affected product for discarding/destruction. This allows us to document your receipt of this recall notice and the status of affected product at
- Quantity in commerce
- 384
- Distribution
- Worldwide distribution - US Nationwide and the countries of Canada, Hong Kong, and Singapore.
- Product codes and lots
- Lot Code: Lot/Serial Number(s) 0004312284 0004314503 0004333110 UDI-DI Each-10889483589151
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1114-2026 | Class I | Ongoing | Voluntary: Firm initiated | ||
| Z-1115-2026 | Class I | Ongoing | Voluntary: Firm initiated | ||
| Z-1116-2026 | Class I | Ongoing | Voluntary: Firm initiated | ||
| Z-1117-2026 | Class I | Ongoing | Voluntary: Firm initiated |