Z-1156-2026Class II
Patient Information Center iX; Software Version Number: 4.5.0
Philips North America LLC / initiated 2025-12-22 / Open, Classified / root cause: software
Philips North America began a recall of Patient Information Center iX; Software Version Number: 4.5.0 on . FDA classified it as Class II and gives the reason as "A recent software (SW) patch modifies Mobile Event Notification filter settings without providing any indication to the user when upgrading the system". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1156-2026
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Philips North America LLC
- Product
- Patient Information Center iX; Software Version Number: 4.5.0
- Product code
- MHX Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)
- Reason for recall
- A recent software (SW) patch modifies Mobile Event Notification filter settings without providing any indication to the user when upgrading the system.
- Root cause tag
- software
- FDA root cause
- Software Design Change
- Action
- On or about December 22, 2025 URGENT Medical Device Correction letters were sent to customers. Actions to be taken: Review the Mobile Notification Event Filters settings and confirm all the events intended to be sent to mobile devices automatically are enabled for Mobile Notification Event Filters. " Pass this notice to all those who need to be aware within your organization or to any organization where affected product(s) have been potentially transferred. Actions planned by Philips to correct the problem " Philips representative will contact you to obtain PIC iX 4.5.1 SW patch for PIC iX systems running software version 4.5.0. If you need any further information, please contact your local Philips representative, or call (800) 722-9377.
- Quantity in commerce
- 146 units (41 US, 105 OUS)
- Distribution
- Worldwide distribution - US Nationwide and the countries of Austria, Belgium, Brazil, Canada, Czech Republic, Dominican Rep, Finland, France, Germany, Greece, Hungary, India, Italy, Japan, Jordan, Netherlands, Norway, Poland, Saudi Arabia, Singapore, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, United Kingdom, Utd.Arab.Emir.
- Product codes and lots
- Model Numbers: (1) 866389, (2) 866390; UDI-DI: (1) 00884838127074, (2) N/A; Serial Numbers: 4Z24-5CKL-P, 0P6C-4TL3-D, 6J58-3LL6-0, 216G-7PGV-E, 0F2U-1BLC-H, 291E-7HJU-A, 1M6W-1UP5-1, 4620-0XNW-5, 4W5R-3FHA-L, 6B73-3ZKP-K, 3E79-49M4-3, 4F6Y-70HU-P, 0128-2WG5-0, 0830-3RLW-W, 097K-7LPC-5, 0D2V-1BLC-8, 120T-17M9-A, 173C-20JM-J, 190B-64MU-U, 1K2J-15LB-P, 1L3T-7KGB-E, 2K06-4ZPR-X, 3916-5DJ6-Z, 3V7H-06MT-E, 412A-2WG5-Y, 4271-2NL7-M, 4Z3L-5PP3-0, 5D0E-3EH3-3, 5E0B-3EH3-D, 5E25-14LG-9, 657X-31N5-L, 6G5U-3FHA-R, 781T-7TJA-B, 7N3U-51MJ-B, 7P31-7NPM-P, 283X-7ZPB-A, 7H77-09GU-Z, 4C59-26K6-P, 4N0L-0FKE-2, 5Y0A-02P6-L, 494H-1CJJ-3, 326G-7PGV-A, 292G-3YNZ-9, 0V2A-6DHP-0, 6354-58M8-C, 4W5B-6JK9-A, 5M7K-0VKT-T, 3401-1FHE-M, 6R30-5BNT-N, 6N2E-7YK1-1, 6R0H-6AMX-T, 107C-62PF-Y, 2X1T-7TJA-H, 1M6G-7PGV-K, 1C4Z-6KKD-H, 0B3Z-2RHR-L, 7H39-3JGK-U, 0M0V-6FKB-5, 793D-41PZ-X, 233G-46MR-V, 782Z-3RLW-D, 5N2B-5YJP-G, 0L5F-0ZNF-J, 4G4N-2LKY-K, 2M48-4XM7-Y, 133W-3UN7-B, 295N-0YG8-Y, 5D7W-00G0-1, 5R78-6CPV-0, 3C79-49M4-9, 3T24-5CKL-A, 6H74-27LY-T, 186W-1UP5-8, 0021-1DNN-P, 0247-7ELE-R, 0D5Y-32PN-A, 0J7W-31N5-2, 0R2W-1VNP-P, 0Y5Z-7FHX-N, 121L-1KG1-M, 1366-7XLV-X, 144W-7RPJ-5, 1472-5WKF-X, 1C19-1MHK-F, 1G0F-69LP-G, 1J5U-5HMV-N, 1P4F-34MB-W, 1P5C-6JK9-D, 212R-6HHJ-P, 221V-7TJA-4, 227N-7JG6-J, 232W-7DL9-G, 2A5X-2PJT-0, 2A75-27LY-3, 2F5J-0JLM-7, 2J19-1MHK-W, 2L16-59N8-M, 2M73-5WKF-U, 2N3V-6YM2-X, 2P6J-7PGV-E, 2Y5M-5AGH-E, 326W-21KC-B, 332M-47P7-Y, 3A29-0CKW-3, 3C0U-7AHH-6, 3C2U-05HF-W, 3E05-4ZPR-U, 3G6H-39G4-Y, 3H1T-22HW-5, 3H5R-5HMV-M, 3H79-09GU-5, 3J2K-2TP9-1, 3W1Y-3CH6-J, 4L7Y-00G0-F, 4M77-1HM0-G, 4N3D-20JM-K, 4R0G-60JV-1, 4V2Z-7NPM-M, 4W08-0UH7-U, 4X6C-1PJ5-J, 4Y08-0UH7-L, 500W-6WNX-A, 594D-0EG2-9, 5U0A-3WPK-C, 5W1G-50P2-N, 5Z6J-3HL1-B, 617W-31N5-F, 637V-7BG9-W, 673T-67ME-B, 686N-3NNV-6, 6D3P-4WMG-A, 6K70-3XNY-W, 6P42-35J4-V, 6U1T-75JN-Y, 771A-2JPY-0, 7F5L-73HZ-3, 7G3A-13JG-L, 7K0P-4AH2-6, 7Z2R-05HF-X, 1J1P-6PKN-T, 4R7H-7LPC-J, 282N-0WPA-X, 5G35-04K8-E, 366L-2XHV-A, 5L7D-2UGW-T, 4N0L-0FKE-2;
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1156-2026 | Class II | Ongoing | Voluntary: Firm initiated |