*HOT AXIOS Stent and Electrocautery-Enhanced Delivery System 8mm x 8mm; Material Number (UPN): M00553530

Boston Scientific Corporation / initiated 2025-12-19 / Open, Classified / root cause: manufacturing process

Boston Scientific Corporation began a recall of *HOT AXIOS Stent and Electrocautery-Enhanced Delivery System 8mm x 8mm; Material Number (UPN): M00553530 on . FDA classified it as Class I and gives the reason as "Increased reports of stent deployment and expansion issues. The most common outcome reasonably foreseeable resulting from difficulty deploying the stent is prolongation of the procedure to exchange the device for a new...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1162-2026
Classification
Class I
Status
Open, Classified
Initiated
Posted
Recalling firm
Boston Scientific Corporation
Product
*HOT AXIOS Stent and Electrocautery-Enhanced Delivery System 8mm x 8mm; Material Number (UPN): M00553530;
Product code
KNS Unit, Electrosurgical, Endoscopic (With Or Without Accessories)
Reason for recall
Increased reports of stent deployment and expansion issues. The most common outcome reasonably foreseeable resulting from difficulty deploying the stent is prolongation of the procedure to exchange the device for a new one. The most serious reasonably foreseeable outcome in cases where the first flange is unable to deploy or expand would be a cascade of events resulting in additional intervention, either endoscopic or surgical, to remove the stent and close the puncture site.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
On December 19, 2025, Urgent Medical Device Removal - Immediate Action Required letters were sent to customers. Actions to be taken: - Immediately stop further distribution or use of any remaining affected devices. Remove them from your facility s inventory and segregate them in a secure location until they can be returned to Boston Scientific. " Immediately post this information in a visible location near the affected products to ensure this information is readily accessible to all handlers and users of the device. " Share this notice with any health care professional within your organization and with any organization where the affected products may have been transferred. " If you are a distributor, this notification must be forwarded to your customers to ensure notification of this product removal is carried out to the end-user level. " If you are a facility that has sent products to another hospital or a facility within your network, ensure this notification is forwarded to them. " Boston Scientific reminds healthcare professionals that AXIOS Stents should only be used as indicated in Instructions For Use.
Quantity in commerce
150 units
Distribution
Worldwide distribution. US Nationwide including Puerto Rico; Algeria, Argentina, Austria, Belgium, Brazil, Canada, CHILE, Colombia, Costa Rica, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Hungary, India, Indonesia, Iran, Ireland, Israel, Italy, JAPAN, Jordan, Kuwait, Lebanon, Luxembourg, Malaysia, Mexico, Morocco, Netherlands, Norway, Oman, Pakistan, Palestine, Panama, Paraguay, PERU, Poland, Portugal, Romania, Saudi Arabia, Serbia, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, UAE, and United Kingdom.
Product codes and lots
Material Number (UPN): M00553530; UDI-DI: 8714729904557; Lot Numbers: 35985705, 35986358, 36147098, 36147222, 36165642, 36176423, 36190575, 36191222, 36194448, 36218745, 36227778, 36227779, 36241500, 36271259, 36280398, 36293880, 36300839, 36301620, 36343103, 36343104, 36364646, 36425887, 36469811, 36471228, 36491879, 36492660, 36506029, 36510220, 36510221, 36513401, 36513402, 36513403, 36517139, 36517220, 36525208, 36525209, 36525210, 36534514, 36534518, 36536259, 36536480, 36536481, 36551000, 36551001, 36642564, 36642565, 36642756, 36645775, 36663654, 36677842, 36677843, 36682759, 36684887, 36684888, 36730665, 36730666, 36747824, 36747825, 36753898, 36753899, 36757493, 36757494, 36761827, 36766370, 36766371, 36783265, 36792467, 36792468, 36792469, 36792470, 36792471, 36799315, 36815090, 36815092, 36832774, 36833683, 36836357, 36844783, 36845422, 36846169, 36846176, 36851617, 36851792, 36853596, 36865505, 36873631, 36873632, 36873906, 36874863, 36904228, 36904229, 36904230, 36907002, 36907003, 36907004, 36907028, 37102533, 37102534, 37102535, 37102536, 37102537, 37459567, 37459568, 37459569, 37459570, 37463133, 37470670, 37498192, 37511243, 37519095, 37519096, 37844029, 37519097, 37520927, 37530320, 37530321, 37530333, 37533308, 37541933, 37571927, 37571928, 37579364, 37579366, 37579367, 37580630, 37585525, 37585526, 37589261, 37589262, 37594668, 37649481, 37649482, 37649483, 37650631, 37694052, 37699011, 37699012, 37708202, 37779125, 37779126, 37780822, 37780823, 37780824, 37823589, 37823590, 37827568, 37827569, 37833165, 37833176, 37834499;

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1162-2026
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1158-2026Class IOngoingVoluntary: Firm initiated
Z-1159-2026Class IOngoingVoluntary: Firm initiated
Z-1160-2026Class IOngoingVoluntary: Firm initiated
Z-1161-2026Class IOngoingVoluntary: Firm initiated
Z-1162-2026Class IOngoingVoluntary: Firm initiated
Z-1163-2026Class IOngoingVoluntary: Firm initiated