Vortex Surgical 1. 23GA Laser Probe Curved Cat No. VS0120.23 2. 25GA Laser Probe Curved Cat No. VS0120.25 3...
Vortex Surgical Inc. began a recall of Vortex Surgical 1. 23GA Laser Probe Curved Cat No. VS0120.23 2. 25GA Laser Probe Curved Cat No. VS0120.25 3. 25GA Laser Probe Flex-Tip Cat No. VS0130.25 4... on . FDA classified it as Class II and gives the reason as "There may be voids located in the seal of Tyvek pouches associated with identified catalog numbers and lots. A compromised sterile barrier has potential of bioburden contamination which could lead to infection". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1245-2026
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Vortex Surgical, Inc.
- Product
- Vortex Surgical 1. 23GA Laser Probe Curved Cat No. VS0120.23 2. 25GA Laser Probe Curved Cat No. VS0120.25 3. 25GA Laser Probe Flex-Tip Cat No. VS0130.25 4. 25GA Laser Probe Illuminated Curved Cat No. VS0125.25B 5. 25GA Laser Probe MaxReach Cat No. VS0140.25 6. 25GA Laser Probe Straight Cat No. VS0110.25
- Product code
- HQB Photocoagulator And Accessories
- Reason for recall
- There may be voids located in the seal of Tyvek pouches associated with identified catalog numbers and lots. A compromised sterile barrier has potential of bioburden contamination which could lead to infection.
- Root cause tag
- manufacturing process
- FDA root cause
- Employee error
- Action
- On December 16, 2025, the firm began notifying customers via "URGENT: MEDICAL DEVICE RECALL" letters. Customers were instructed to discontinue use of affected product and to return product for Vortex Surgical to evaluate the seals; use UPS account 23623Y to ship unused products to Vortex Surgical Inc for replacement or refund at: Attn: Andrew Ritts Recall Tyvek Seal, Vortex Surgical Inc., 4 Research Park Dr., Suite 124, St. Charles, MO 63304; complete and return the Acknowledgement and Receipt Form via mail or email to: [email protected]. Actions have been implemented to improve training of individuals associated with sealing and inspecting seals to prevent reoccurrence. The sealer associated with the issue is being updated to include a guard to prevent reoccurrence. If you have any questions, contact the Director of Regulatory Affairs at 635-778-4350 or email to: [email protected] from 8am to 5pm.
- Quantity in commerce
- 14,789 (8651 US; 6138 OUS)
- Distribution
- Worldwide Distribution: US (nationwide) to states of: AL, AR, AZ, CA, CT, FL, HI, IA, IL, IN, KY, MA, MI, MN, MS, NC, NE, NJ, NM, NY, OH, OR, PA, RI, SC, SD, TN, TX, WA, WI, VA; and OUS (foreign) to countries of: Australia, Chile, China, France, Germany, India, Ireland, Italy, Japan, Korea, Mexico, Netherlands, Philippines, Poland, Puerto Rico, Spain, Switzerland, Taiwan, United Kingdom, Vietnam.
- Product codes and lots
- 1. 23GA Laser Probe Curved Pouch UDI 810123480036 Box UDI 810123480180 Lot 2411027 2. 25GA Laser Probe Curved Pouch UDI 810123480043 Box UDI 810123480197 Lot 2411026 3. 25GA Laser Probe Flex-Tip Pouch UDI 810123480111 Box UDI 810123480265 Lot 2411024 2411031 4. 25GA Laser Probe Illuminated Curved Pouch UDI 810123480098 Box UDI 810123480241 Lot 2411025 5. 25GA Laser Probe MaxReach Pouch UDI 810123480135 Box UDI 810123480289 Lot 2410029 2411023 6. 25GA Laser Probe Straight Pouch UDI 810123480012 Box UDI 810123480166 Lot 2411030
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1244-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1245-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1246-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1247-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1248-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1249-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1250-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1251-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1252-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1253-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1254-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1255-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1256-2026 | Class II | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.