MONARCH Platform, Bronchoscopy, Catalog Numbers: MON-000005-01 , MON-000005-01R , MON-000006...

Auris Health, Inc / initiated 2026-01-21 / Open, Classified / root cause: software

Auris Health, Inc began a recall of MONARCH Platform, Bronchoscopy, Catalog Numbers: MON-000005-01 , MON-000005-01R , MON-000006 , MON-000006-RFB, MON-000008 with software on . FDA classified it as Class II and gives the reason as "Device for bronchoscopic visualization, patient airway access has software issue: if application restarts after patient-side selection, prior to bronchoscope loading, application will re-initialize to patient-left...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.

Recall record

Recall number
Z-1383-2026
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Auris Health, Inc
Product
MONARCH Platform, Bronchoscopy, Catalog Numbers: MON-000005-01 , MON-000005-01R , MON-000006 , MON-000006-RFB, MON-000008 with software
Product code
EOQ Bronchoscope (Flexible Or Rigid)
Reason for recall
Device for bronchoscopic visualization, patient airway access has software issue: if application restarts after patient-side selection, prior to bronchoscope loading, application will re-initialize to patient-left position; and if right position was previously selected, this may lead to unanticipated robotic arm positioning toward patient-left, may result in contact with patient/endotracheal tube.
Root cause tag
software
FDA root cause
Software design
Action
On 1/21/2026, correction notices were emailed to customers informing them of the following: Should a MONARCH application restart occur during procedure setup, please clear the fault message and then perform a full system restart. A full system restart will prevent this issue from occurring as it will require the user to perform patient-side selection and alignment again. Once the software update to address the issue is available, the firm will contact customers to schedule installation. Complete and return the acknowledgement form via email to [email protected] This notice needs to be passed on to all those who need to be aware within your organization or to any organization where the potential affected devices have been transferred. Please transfer this notice to other organizations on which this action has an impact. Maintain awareness on this notice and resulting action for an appropriate period to ensure effectiveness of the corrective actions. If you have questions contact the firm at +1 (800) 434-0032 (toll-free within the United States), +1 (650) 264-6000 (Worldwide) or via email at [email protected] For Medical Information requests, visit firm's website at https://www.jnjmedtech.com/mir.
Quantity in commerce
173
Distribution
Worldwide - US Nationwide distribution in the states of MN, FL, NJ, NY, VA, WI, AZ, AL, MS, LA, MI, MA, MT, PA, WV, GA, DE, TX, CA, OH, IN, SC, CT, ND, MO, DC, IL, NE, ID, WA, NC, TN, OR, CO, SD, NM, VT, MD and the countries of Hong Kong, China, Canada.
Product codes and lots
Software versions: 2.2.2, 2.2.3, 2.2.4, 2.2.5, and 4.1.1. UDI-DI: 10810068810803, 10810068810988. Serial Numbers: 110024, 120088, 120057, 120046, 120058, 120015, 120160, 120068, 120081, 120056, 110028, 120128, 120017, 120089, 120075, 110016, 120031, 120169, 120113, 120168, 120055, 120014, 120060, 120174, 120195, 120026, 110038, 120124, 120194, 110027, 110035, 120123, 120091, 120007, 110022, 120006, 120152, 120004, 120120, 120041, 120106, 110034, 120126, 120082, 120151, 120051, 120020, 120153, 120193, 120170, 120108, 120037, 120074, 120199, 120130, 120095, 110026, 120096, 120009, 120111, 120022, 120054, 120163, 110045, 120027, 110039, 120090, 120141, 120072, 120155, 110049, 120143, 110047, 120125, 120052, 110052, 120100, 120050, 120098, 120045, 120131, 110020, 110030, 110032, 110021, 120013, 120105, 120136, 120192, 110017, 120029, 120139, 110031, 120137, 120053, 110029, 120028, 120034, 120085, 120019, 120049, 120209, 120144, 120161, 120001, 110033, 120145, 120127, 110046, 110043, 120011, 120070, 120076, 120134, 120117, 110025, 120077, 120149, 120191, 120102, 120002, 120156, 120010, 120171, 120003, 120023, 120146, 120188, 120043, 120071, 120042, 120196, 110051, 120065, 120172, 120116, 120150, 120066, 120166, 120033, 110013, 120079, 120008, 120086, 120032, 120073, 1806P1, 120197, 120080, 120078, 14, 120012, 110041, 120084, 120219, 110036, 120181, 120083, 120067, 120021, 120216, 120217, 120218, 120069, 120204, 120207, 120215, 120237, 120208, 120213, 120164, 120132, 120182

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1383-2026
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1383-2026Class IIOngoingVoluntary: Firm initiated