Medline ReNewal Reprocessed ViewFlex Xtra ICE Catheters: ViewFlex Xtra ICE Catheter 9F x 90cm D087031RH...
Medline Industries, LP began a recall of Medline ReNewal Reprocessed ViewFlex Xtra ICE Catheters: ViewFlex Xtra ICE Catheter 9F x 90cm D087031RH 20197344019554 on . FDA classified it as Class I and gives the reason as "These lots of reprocessed catheters may contain small particles of residual material. If affected devices are used, there is a risk of inflammatory response, systemic infection, or granulomatous reaction or thrombus...". As of , the recall status on record is Open, Classified.
Recall record
- Recall number
- Z-1323-2026
- Classification
- Class I
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- MEDLINE INDUSTRIES, LP - Northfield
- Product
- Medline ReNewal Reprocessed ViewFlex Xtra ICE Catheters: ViewFlex Xtra ICE Catheter 9F x 90cm D087031RH 20197344019554
- Product code
- OWQ Reprocessed Intravascular Ultrasound Catheter
- Reason for recall
- These lots of reprocessed catheters may contain small particles of residual material. If affected devices are used, there is a risk of inflammatory response, systemic infection, or granulomatous reaction or thrombus reaction which may lead to cerebral or pulmonary embolism, or deep vein thrombosis. Expansion of recall Z-2610/2614-2025
- FDA root cause
- Under Investigation by firm
- Action
- On December 22, 2025, the firm issued Urgent Medical Device Recall letters as a follow-up to an earlier action in 2025 for the same issue, but newly identified item numbers and lots of previously identified item numbers. Notably, the products associated with the new/updated action are to be destroyed by the customer on-site for credit, and not returned to the manufacturer. ***Updated 5/18/26*** The firm issued an expanded notification on 5/15/26 which expanded the action to include on additional lot of product. These devices were not expired at the time of the original notification but have since expired.
- Quantity in commerce
- 650
- Distribution
- Distribution US nationwide.
- Product codes and lots
- UDI-DI (case) 20197344019554¿ (ea) 10197344019557 Lots EP250324 EP250416 EP250512 EP250521 EP250609 EP250630 EP250710 EP250724
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1319-2026 | Class I | Ongoing | Voluntary: Firm initiated | ||
| Z-1320-2026 | Class I | Ongoing | Voluntary: Firm initiated | ||
| Z-1321-2026 | Class I | Ongoing | Voluntary: Firm initiated | ||
| Z-1322-2026 | Class I | Ongoing | Voluntary: Firm initiated | ||
| Z-1323-2026 | Class I | Ongoing | Voluntary: Firm initiated | ||
| Z-1324-2026 | Class I | Ongoing | Voluntary: Firm initiated | ||
| Z-1325-2026 | Class I | Ongoing | Voluntary: Firm initiated |