Z-1220-2026Class II

BEAR Implant. Model Number: 1000

Miach Orthopaedics / initiated 2025-12-31 / Open, Classified / root cause: manufacturing process

Miach Orthopaedics began a recall of BEAR Implant. Model Number: 1000 on . FDA classified it as Class II and gives the reason as "Incorrect expiration date on label that extends beyond the product's documented/approved shelf life. With an incorrect expiration date, the UDI will also be incorrect". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1220-2026
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Miach Orthopaedics
Product
BEAR Implant. Model Number: 1000.
Product code
QNI Resorbable Implant For Anterior Cruciate Ligament (Acl) Repair
Reason for recall
Incorrect expiration date on label that extends beyond the product's documented/approved shelf life. With an incorrect expiration date, the UDI will also be incorrect.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Notification to Sales Representatives and consignees began on about 12/31/2025 via email and letter titled URGENT MEDICAL DEVICE RECALL NOTIFICATION. Consignees were instructed to identify and quarantine all unused inventory of the affected lot, discontinue distribution of the affected lot, return all unused affected product, and complete and return the provided Recall Acknowledgement Form. Additionally, consignees were requested to notify customers if the affected lot was further distributed.
Quantity in commerce
284 units
Distribution
US Nationwide distribution in the states of AK, CA, CO, CT, FL, GA, IL, MA, MD, ME, MI, MN, MO, NC, NJ, NY, OH, OR, TX, VA, WA, WI.
Product codes and lots
Model Number: 1000. UDI Number: (01)00860002987804(17)260228(10)7011098. Lot Number: 7011098. Expiration Date: 02/28/2026.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1220-2026
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1220-2026Class IIOngoingVoluntary: Firm initiated