Z-1456-2026Class II
Olympus Inner Sheath. Model No. A2642. for endoscopic diagnosis and treatment in urological applications
OLYMPUS Corporation of the Americas / initiated 2025-11-11 / Open, Classified
Olympus Corporation of the Americas began a recall of Olympus Inner Sheath. Model No. A2642. for endoscopic diagnosis and treatment in urological applications on . FDA classified it as Class II and gives the reason as "Complaints of the ceramic tip of the resection sheath breaking have been received". As of , the recall status on record is Open, Classified.
Recall record
- Recall number
- Z-1456-2026
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- OLYMPUS Corporation of the Americas
- Product
- Olympus Inner Sheath. Model No. A2642. for endoscopic diagnosis and treatment in urological applications.
- Product code
- FJL Resectoscope
- Reason for recall
- Complaints of the ceramic tip of the resection sheath breaking have been received.
- FDA root cause
- Under Investigation by firm
- Action
- An URGENT MEDICAL DEVICE CORRECTIVE ACTION notice was sent to consignees 1/6/2026. Consignees are instructed to ensure all personnel are familiar with the contents of the notification. Users are to adhere to the Inspection and Testing and Caution sections of the IFU for both the resection/inner sheath, as well as the Inspection Before Each Use section in the Resectoscope System Manual. Additional instructions include having an extra resection sheath and inspect during and after procedure for device fragments. Consignees are to retain a copy of the notification with the IFU. Consignees with any questions may contact Cynthia Ow at [email protected] or by phone at 947-999-3203.
- Quantity in commerce
- 0 units
- Distribution
- US Nationwide Distribution.
- Product codes and lots
- Model No. A2642; UDI: 04042761004084; All Lots.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1438-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1439-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1440-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1441-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1442-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1443-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1444-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1445-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1446-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1447-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1448-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1449-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1450-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1451-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1452-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1453-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1454-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1455-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1456-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1457-2026 | Class II | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.