Medline Industries IV Administration and Extension Sets: 1) DBD-IV ADMIN SET,60DRP,1NEEDLE-FREE VALV, Model...
Medline Industries, LP began a recall of Medline Industries IV Administration and Extension Sets: 1) DBD-IV ADMIN SET,60DRP,1NEEDLE-FREE VALV, Model Number: DYNDTB0529; 2) IV ADMIN SET, 10DRP, 1NF... on . FDA classified it as Class II and gives the reason as "Under circumstances involving exposure to particular chemicals and mechanical forces, the male luer connectors for the specified IV Administration and Extension sets may leak, crack, and/or break during use, which may...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1366-2026
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- MEDLINE INDUSTRIES, LP - Northfield
- Product
- Medline Industries IV Administration and Extension Sets: 1) DBD-IV ADMIN SET,60DRP,1NEEDLE-FREE VALV, Model Number: DYNDTB0529; 2) IV ADMIN SET, 10DRP, 1NF, 92, Model Number: DYNDTB1029; 3) DBD-IVEXTSET2.4ML2 NEEDLE-FREEVLVE,35, Model Number: DYNDTB1516; 4) DBD-IV ADMIN SET, 15DRP, 1 INJECT PORT,, Model Number: DYNDTB1530; 5) DBD-IV ADMIN SET, 15DRP, ROLLER CLAMP, 6, Model Number: DYNDTB1539; 6) DBD-IV ADMIN SET, 15DRP, 3 NF, 109, Model Number: DYNDTB1545; 7) IV ADMIN SET, 20DRP, 3 NF, 109, Model Number: DYNDTB2045; 8) DBD-IVADMINSET60DRP2NEEDLE-FREEVLVE96, Model Number: DYNDTB0512; 9) DBD-IVADMINSET,60DRP3 NEEDLE-FREE VLVE11, Model Number: DYNDTB0555; 10) IV ADMIN SET, 10DRP, NF,INJ, 96, Model Number: DYNDTB1022; 11) DBD-IV ADMIN SET, 15DRP, 2NF, 96, Model Number: DYNDTB1512; 12) DBD-IV ADMIN SET, 15DRP, 1NF, 92, Model Number: DYNDTB1529; 13) DBD-IV ADMIN SET, 15DRP, ROLLER CLAMP, 9, Model Number: DYNDTB1537; 14) DBD-IV ADMIN SET, 15DRP, SECOND, 40, Model Number: DYNDTB1540; 15) IV ADMIN SET, 20DRP, 1NF, 92, Model Number: DYNDTB2029; 16) DBD-IV EXT SET, 6.5, 1.1ML, M LUERLOCK,, Model Number: DYNDTB5077; 17) IV ADMIN SET, 10DRP, 3 NF, 109, Model Number: DYNDTB1045; 18) DBD-IVADMISET15DRP4NEEDLEFREE VLVESTPCK1, Model Number: DYNDTB1545D; 19) IV ADMIN SET15DRPFLWCTRL1NEEDLEFREE VA, Model Number: DYNDTB2524; 20) IV ADMIN SET, 10DRP, 2NF, 96, Model Number: DYNDTB1012; 21) IV ADMIN SET, 20DRP, 2NF, 96, Model Number: DYNDTB2012; 22) IV ADMIN SET, 20DRP, NF,INJ, 96, Model Number: DYNDTB2022; 23) DBD-IV EXT SET,SMBORE EXT,8,M LUERLOCK,1, Model Number: DYNDTB5081
- Product code
- FPA Set, Administration, Intravascular
- Reason for recall
- Under circumstances involving exposure to particular chemicals and mechanical forces, the male luer connectors for the specified IV Administration and Extension sets may leak, crack, and/or break during use, which may result in a delay in patient treatment, blood loss, infection, an air embolism, and/or underdosage of medication.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- An URGENT MEDICAL DEVICE RECALL notification letter dated 12/31/25 was sent to customers. REQUIRED ACTION: 1. Immediately check your stock for the affected item number(s) and the affected lot number(s) listed within the recall portal. Quarantine all affected product. Upon completion of the below actions, please destroy affected product. 2. Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected product in inventory, please complete and submit the response form. Website link: https://recalls.medline.com Recall Reference #: R-25-262 Recall Code: 3. Your account will receive credit once the response form is submitted. 4. If you are a distributor, or have resold or transferred this product to another company or individual, you are required by FDA regulations to notify them of this recall communication. Please request that customers destroy any affected product. You should include your customers quantities on your response form. If you have any questions, contact the Recall Department at 866-359-1704 or [email protected].
- Quantity in commerce
- 926,227 units
- Distribution
- Worldwide distribution - US Nationwide and the country of Bahamas.
- Product codes and lots
- 1) DYNDTB0529, UDI-DI: 10197344105663(each), 20197344105660(case), Lot Number: All Lot Numbers; 2) DYNDTB1029, UDI-DI: 10197344076956(each), 20197344076953(case), Lot Number: All Lot Numbers; 3) DYNDTB1516, UDI-DI: 10197344105687(each), 20197344105684(case), Lot Number: All Lot Numbers; 4) DYNDTB1530, UDI-DI: 10197344105694(each), 20197344105691(case), Lot Number: All Lot Numbers; 5) DYNDTB1539, UDI-DI: 10197344008254(each), 20197344008251(case), Lot Number: All Lot Numbers; 6) DYNDTB1545, UDI-DI: 10197344098644(each), 20197344098641(case), Lot Number: All Lot Numbers; 7) DYNDTB2045, UDI-DI: 10197344077007(each), 20197344077004(case), Lot Number: All Lot Numbers; 8) DYNDTB0512, UDI-DI: 10197344105656(each), 20197344105653(case), Lot Number: All Lot Numbers; 9) DYNDTB0555, UDI-DI: 10197344105670(each), 20197344105677(case), Lot Number: All Lot Numbers; 10) DYNDTB1022, UDI-DI: 10197344076949(each), 20197344076946(case), Lot Number: All Lot Numbers; 11) DYNDTB1512, UDI-DI: 10197344098613(each), 20197344098610(case), Lot Number: All Lot Numbers; 12) DYNDTB1529, UDI-DI: 10197344098620(each), 20197344098627(case), Lot Number: All Lot Numbers; 13) DYNDTB1537, UDI-DI: 10197344008247(each), 20197344008244(case), Lot Number: All Lot Numbers; 14) DYNDTB1540, UDI-DI: 10197344098637(each), 20197344098634(case), Lot Number: All Lot Numbers; 15) DYNDTB2029, UDI-DI: 10197344076994(each), 20197344076991(case), Lot Number: All Lot Numbers; 16) DYNDTB5077, UDI-DI: 10197344098651(each), 20197344098658(case), Lot Number: All Lot Numbers; 17) DYNDTB1045, UDI-DI: 10197344076963(each), 20197344076960(case), Lot Number: All Lot Numbers; 18) DYNDTB1545D, UDI-DI: 10197344105700(each), 20197344105707(case), Lot Number: All Lot Numbers; 19) DYNDTB2524, UDI-DI: 10197344105717(each), 20197344105714(case), Lot Number: All Lot Numbers; 20) DYNDTB1012, UDI-DI: 10197344076932(each), 20197344076939(case), Lot Number: All Lot Numbers; 21) DYNDTB2012, UDI-DI: 10197344076970(each), 20197344076977(case), Lot Number: All Lot Numbers; 22) DYNDTB2022, UDI-DI: 10197344076987(each), 20197344076984(case), Lot Number: All Lot Numbers; 23) DYNDTB5081, UDI-DI: 10197344098668(each), 20197344098665(case), Lot Number: All Lot Numbers