Z-1485-2026Class II

Brand Name: AESCULAP Product Name: MINOP TROCAR 150MM 4 WKING CHANNEL6.0MM Model/Catalog Number: FF399R...

Aesculap Inc / initiated 2026-01-15 / Open, Classified

Aesculap Inc began a recall of Brand Name: AESCULAP Product Name: MINOP TROCAR 150MM 4 WKING CHANNEL6.0MM Model/Catalog Number: FF399R Software Version: N/A Product Description: MINOP TROCAR... on . FDA classified it as Class II and gives the reason as "There is the potential for the length of the trocar shaft to be too long". As of , the recall status on record is Open, Classified.

Recall record

Recall number
Z-1485-2026
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Aesculap Inc
Product
Brand Name: AESCULAP Product Name: MINOP TROCAR 150MM 4 WKING CHANNEL6.0MM Model/Catalog Number: FF399R Software Version: N/A Product Description: MINOP TROCAR 150MM 4 WKING CHANNEL6.0MM Component: N/A
Product code
GWG Endoscope, Neurological
Reason for recall
There is the potential for the length of the trocar shaft to be too long.
FDA root cause
Under Investigation by firm
Action
on January 28, 2026 URGENT MEDICAL DEVICE CORRECTION NOTIFICATION letters were sent to customers. Actions to be taken: - Review the attachment Aesculap AG Urgent Field Safety Notice in its entirety. - If it is determined that a portion of the visual field is obscured, the manufacturer recommends rotating the device to compensate for the restricted view (refer to the attached Urgent Field Safety Notice from the manufacturer). Note: Devices are not being removed from the market, this is a field notification. - Return the completed Urgent Medical Device Correction Acknowledgement Form to Aesculap, Inc. Quality Assurance department by faxing the form to (610) 849-1197 or e-mail to [email protected] within two (2) weeks of receipt. Action planned by Aesculap AG to correct the problem: To address this issue, the manufacturer is implementing additional inspections of the optical system and introducing clearly defined acceptance criteria, along with other corrective actions. Should there be any questions, comments, or concerns regarding the product you have received, please feel free to contact (844-903-6417) or [email protected]
Quantity in commerce
126 units
Distribution
US Nationwide distribution in the states of AZ, AR, CA, CO, FL, GA, HI, IA, ID, IL, KS, LA, MA, MD, MI, MO, MS, NC, NE, NJ, NY, OH, OK, OR, PA, RI, SC, SD, TX, UT, WA.
Product codes and lots
Mode No. FF399R; UDI-DI 04038653065547; Serial Numbers: 4595, 4619, 4657, 4764, 4592, 4596, 4598, 4601, 4606, 4607, 4608, 4612, 4616, 4618, 4620, 4621, 4623, 4659, 4661, 4662, 4667, 4670, 4673, 4674, 4688, 4689, 4746, 4748, 4757, 4758, 4760, 4761, 4762, 4767, 4768, 4771, 4773, 4774, 4779, 4780, 4784, 4786, 4789, 4791, 4792, 4795, 4799, 4811, 4812, 4818, 4819, 4821, 4822, 4824, 4825, 4826, 4827, 4836, 4838, 4841, 4842, 4843, 4854, 4856, 4858, 4864, 4868, 4869, 4880, 4882, 4886, 4890, 4891, 4892, 4893, 4899, 4901, 4902, 4907, 4914, 4921, 4927, 4940, 4951, 4959, 4967, 4968, 4970, 4971, 4972, 4973, 4974, 4975, 4976, 4990, 4994, 4995, 5000, 5004, 5015, 5024, 5035, 5036, 5042, 5057, 5058, 5059, 5062, 5063, 5064, 5072, 5076, 5081, 5084, 5106, 5112, 5115, 5117, 5118, 5119, 5121, 5123, 5165, 5166, 5168, 5172;

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1485-2026
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1485-2026Class IIOngoingVoluntary: Firm initiated