Z-1370-2026Class II

Medica Capillary Tubes; EasyBloodGas/EasyStat Capillary Tube Kit

Medica Corp. / initiated 2025-12-31 / Open, Classified / root cause: manufacturing process

Medica Corporation began a recall of Medica Capillary Tubes; EasyBloodGas/EasyStat Capillary Tube Kit on . FDA classified it as Class II and gives the reason as "A systematic positive bias of up to 20% was identified across the reportable potassium range, resulting in false elevation results of patient samples when using Medica s capillary tubes". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1370-2026
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Medica Corp.
Product
Medica Capillary Tubes; EasyBloodGas/EasyStat Capillary Tube Kit.
Product code
CEM Electrode, Ion Specific, Potassium
Reason for recall
A systematic positive bias of up to 20% was identified across the reportable potassium range, resulting in false elevation results of patient samples when using Medica s capillary tubes.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
An URGENT FIELD SAFETY NOTICE was sent to consignees on 01/09/25 informing them that potassium results obtained using affected devices in capillary mode may be falsely elevated. The FSN instructed consignees not to report potassium results obtained in capillary mode using affected lots on the Easy Stat Analyzer. The FSN is to be forwarded to end users and to where product was further distributed. Consignees are to contact their local representative with any questions.
Quantity in commerce
1,578 units
Distribution
International distribution to the countries of Russia, Bulgaria, Kazakhstan, Ukraine, Pakistan, Belarus.
Product codes and lots
Model No. 7303, UDI-DI: 00840095607575, Lot No. 25XXX.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1370-2026
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1370-2026Class IIOngoingVoluntary: Firm initiated