Medica Capillary Tubes; EasyBloodGas/EasyStat Capillary Tube Kit
Medica Corporation began a recall of Medica Capillary Tubes; EasyBloodGas/EasyStat Capillary Tube Kit on . FDA classified it as Class II and gives the reason as "A systematic positive bias of up to 20% was identified across the reportable potassium range, resulting in false elevation results of patient samples when using Medica s capillary tubes". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1370-2026
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Medica Corp.
- Product
- Medica Capillary Tubes; EasyBloodGas/EasyStat Capillary Tube Kit.
- Product code
- CEM Electrode, Ion Specific, Potassium
- Reason for recall
- A systematic positive bias of up to 20% was identified across the reportable potassium range, resulting in false elevation results of patient samples when using Medica s capillary tubes.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- An URGENT FIELD SAFETY NOTICE was sent to consignees on 01/09/25 informing them that potassium results obtained using affected devices in capillary mode may be falsely elevated. The FSN instructed consignees not to report potassium results obtained in capillary mode using affected lots on the Easy Stat Analyzer. The FSN is to be forwarded to end users and to where product was further distributed. Consignees are to contact their local representative with any questions.
- Quantity in commerce
- 1,578 units
- Distribution
- International distribution to the countries of Russia, Bulgaria, Kazakhstan, Ukraine, Pakistan, Belarus.
- Product codes and lots
- Model No. 7303, UDI-DI: 00840095607575, Lot No. 25XXX.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1370-2026 | Class II | Ongoing | Voluntary: Firm initiated |