Medline Kits containing Tego Connectors Medline DIALYSIS ON/OFF KIT SKU EBSI1746
Medline Industries, LP began a recall of Medline Kits containing Tego Connectors Medline DIALYSIS ON/OFF KIT SKU EBSI1746 on . FDA classified it as Class II and gives the reason as "Affected lots may exhibit issues with the silicone seal on Tego Connectors included in kits. Issues include silicone seal doming and silicone seal tearing. These issues may result in an occluded fluid path identified by...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1464-2026
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- MEDLINE INDUSTRIES, LP - Northfield
- Product
- Medline Kits containing Tego Connectors Medline DIALYSIS ON/OFF KIT SKU EBSI1746
- Product code
- FKG Tray, Start/Stop (Including Contents), Dialysis
- Reason for recall
- Affected lots may exhibit issues with the silicone seal on Tego Connectors included in kits. Issues include silicone seal doming and silicone seal tearing. These issues may result in an occluded fluid path identified by the inability of the user to inject or withdraw blood product with a syringe. Damaged seals may also result in delay in therapy or fluid leakage, interruption in therapy, exposure to biological contaminants or air infused into the body.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- On January 8, 2026, the firm notified affected customers of the product issue through email and first class mail. Customers were instructed to request stickers to over-label affected convenience kits with a warning label indicating to remove the affected component and replace it with product from supply. No kits are to returned.
- Quantity in commerce
- 2020
- Distribution
- US Nationwide distribution.
- Product codes and lots
- UDI-DI (ea) 10653160373582 UDI-DI (case) 00653160373585 Kit lots 2024051380 2024052080 2024052180 2024052280 2024071580 2025061790 2025072990
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1463-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1464-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1465-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1466-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1467-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1468-2026 | Class II | Ongoing | Voluntary: Firm initiated |