Z-1597-2026Class II

Olecranonplate with Angular Stability with the below descriptions and corresponding article numbers. 1...

I.T.S. GmbH / initiated 2026-01-23 / Open, Classified / root cause: labeling and UDI

I.T.S. GmbH began a recall of Olecranonplate with Angular Stability with the below descriptions and corresponding article numbers. 1. Olecranon Hook Plate; Article Numbers: 21112-7... on . FDA classified it as Class II and gives the reason as "Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU)". As of , the recall status on record is Open, Classified, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-1597-2026
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
I.T.S. GmbH
Product
Olecranonplate with Angular Stability with the below descriptions and corresponding article numbers. 1. Olecranon Hook Plate; Article Numbers: 21112-7, 21114-5, 21113-9. 2. Olecranon Plate; Article Numbers: 21111-6, 21111-8, 21111-12.
Product code
HRS Plate, Fixation, Bone
Reason for recall
Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).
Root cause tag
labeling and UDI
FDA root cause
No Marketing Application
Action
ITS initially notified its importer/distributor on about 01/23/2026 via email. Instructions were to cease distribution, and further information will be provided. An additional notification was sent to distributors and consignees on about 02/04/2026 via email. Instructions included to ensure any IFU different to the provided one is removed from systems and/or destroyed, notify clinical staff managing patients with implants, contact I.T.S. GmbH for any questions or to discuss proper scanning conditions for implanted devices in the interim while they work on labeling and IFU updates for the devices, and arrange for the return of specific originally packaged items - base plates and screws in surgery trays. Other outlined products can continue to be used upon confirmation from ITS. Consignees were also instructed to complete and return the response form. Distributors were instructed to continue to immediately halt product distribution from the U.S. office to sales groups and hospitals, remove the outdated paper IFU from packaging and add the eIFU symbol to product labeling to implement the updated IFU provided electronically - see annex "F-7.5.5-3-US_IFU_non-sterile_Rev00_DRAFT". Paper IFUs will be made available upon request. They were also requested to assist in notification of hospitals.
Quantity in commerce
70 units
Distribution
US Nationwide distribution in the states of AL, AZ, BVI, CA, CO, FL, GA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NE, NJ, NM, NV, OH, OK, PA, SC, TX, USVI, WV, W.
Product codes and lots
1 Article number (Lot Numbers): 21112-7 (643/0549), 21114-5 (643/032301, 643/042303, 643/042304, 643/0618, 643/102211, 643/112304), 21113-9 (643/072320). 2. Article number (Lot Numbers): 21111-6 (643/032306, 643/072319, 643/112201), 21111-8 (643/032318), 21111-12 (643/032305).

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1597-2026
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1592-2026Class IIOngoingVoluntary: Firm initiated
Z-1593-2026Class IIOngoingVoluntary: Firm initiated
Z-1594-2026Class IIOngoingVoluntary: Firm initiated
Z-1595-2026Class IIOngoingVoluntary: Firm initiated
Z-1596-2026Class IIOngoingVoluntary: Firm initiated
Z-1597-2026Class IIOngoingVoluntary: Firm initiated
Z-1598-2026Class IIOngoingVoluntary: Firm initiated
Z-1599-2026Class IIOngoingVoluntary: Firm initiated
Z-1600-2026Class IIOngoingVoluntary: Firm initiated
Z-1601-2026Class IIOngoingVoluntary: Firm initiated
Z-1602-2026Class IIOngoingVoluntary: Firm initiated
Z-1603-2026Class IIOngoingVoluntary: Firm initiated
Z-1604-2026Class IIOngoingVoluntary: Firm initiated
Z-1605-2026Class IIOngoingVoluntary: Firm initiated
Z-1606-2026Class IIOngoingVoluntary: Firm initiated
Z-1607-2026Class IIOngoingVoluntary: Firm initiated
Z-1608-2026Class IIOngoingVoluntary: Firm initiated