Z-1550-2026Class II
Brand Name: APTUS Product Name: 2.8 TriLock Screw 16mm, HD7, 1/Pkg Model/Catalog Number: A-5850.16/1 Software...
Medartis AG / initiated 2026-01-09 / Open, Classified / root cause: labeling and UDI
Medartis AG began a recall of Brand Name: APTUS Product Name: 2.8 TriLock Screw 16mm, HD7, 1/Pkg Model/Catalog Number: A-5850.16/1 Software Version: Not applicable. Product Description: The... on . FDA classified it as Class II and gives the reason as "Mix up between specified lots of 2.5mm and 2.8mm outer diameter screws". As of , the recall status on record is Open, Classified, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-1550-2026
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Medartis AG
- Product
- Brand Name: APTUS Product Name: 2.8 TriLock Screw 16mm, HD7, 1/Pkg Model/Catalog Number: A-5850.16/1 Software Version: Not applicable. Product Description: The APTUS fixation systems are used for fractures, osteotomies and arthrodesis of the hand, forearm, shoulder and foot. Component: Not applicable.
- Product code
- HWC Screw, Fixation, Bone
- Reason for recall
- Mix up between specified lots of 2.5mm and 2.8mm outer diameter screws.
- Root cause tag
- labeling and UDI
- FDA root cause
- Labeling mix-ups
- Action
- Firm began notifying affected customers on January 13, 2026 via email. Customers were instructed to return impacted products to Medartis Inc.
- Quantity in commerce
- 126
- Distribution
- Worldwide - US Nationwide distribution in the states of Indiana and Pennsylvania and the countries of Austria, Belgium, Switzerland, Czech Republic, Germany, Spain, Finland, France, Israel, Italy, Japan, and Poland.
- Product codes and lots
- Lot Code: Version or Model: A-5850.16/1 Device Description: 2.8 TriLock Screw 16mm, HD7, 1/Pkg Lot: 25467933
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1549-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1550-2026 | Class II | Ongoing | Voluntary: Firm initiated |