Z-1484-2026Class II
Atellica CH Urine Albumin (UAlb). Material Number: 11537225
Siemens Healthcare Diagnostics Inc. / initiated 2026-01-20 / Open, Classified
Siemens Healthcare Diagnostics, Inc. began a recall of Atellica CH Urine Albumin (UAlb). Material Number: 11537225 on . FDA classified it as Class II and gives the reason as "Falsely depressed UAlb patient results may occur. Affected samples with any urine albumin concentration above the UAlb measuring interval >40.0 mg/dL may report values between 6.2 mg/dL and 40.0 mg/dL". As of , the recall status on record is Open, Classified.
Recall record
- Recall number
- Z-1484-2026
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Product
- Atellica CH Urine Albumin (UAlb). Material Number: 11537225
- Product code
- JIR Indicator Method, Protein Or Albumin (Urinary, Non-Quant.)
- Reason for recall
- Falsely depressed UAlb patient results may occur. Affected samples with any urine albumin concentration above the UAlb measuring interval >40.0 mg/dL may report values between 6.2 mg/dL and 40.0 mg/dL.
- FDA root cause
- Under Investigation by firm
- Action
- Siemens Healthineers notified consignees on about 01/20/2026 via mailed FedEx letter titled Urgent Medical Device Correction. Consignees were informed that as temporary mitigation, for any sample with a UAlb result above 5.0 mg/dL (50 mg/L), reanalyze the sample with a x10 dilution. If the diluted result is consistent with the initial result, either result may be reported. If the diluted result is inconsistent with the initial result, the diluted result should be reported. If you prefer to automate this dilution, please refer to your internal laboratory procedures. Consignees were instructed to review the notification with their Medical Director to determine the appropriate course of action, including evaluation of any previously generated results, if applicable, and complete and return the provided Field Correction Effectiveness Check Form. A follow-up communication will be provided when Customer Actions are no longer required.
- Quantity in commerce
- 4,885 units
- Distribution
- Worldwide distribution - US Nationwide and the countries of Australia, Austria, Belgium, Brazil, Canada, Chile, Curacao,¿St¿Eus, Czech¿Republic, Denmark, Finland, France, Germany, Greece, Hong Kong, Ireland, Israel, Italy, Korea, Latvia, Malaysia, Netherlands, New Zealand, Norway, Portugal, Romania, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, United¿Kingdom.
- Product codes and lots
- Material Number: 11537225. UDI-DI: 00630414611099. Lot Numbers: All lot numbers.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1484-2026 | Class II | Ongoing | Voluntary: Firm initiated |