Z-1611-2026Class II
Olympus Endoscope Air/Water Valve - MAJ-1444. Model Number: MAJ-1444
OLYMPUS Corporation of the Americas / initiated 2026-02-12 / Open, Classified
Olympus Corporation of the Americas began a recall of Olympus Endoscope Air/Water Valve - MAJ-1444. Model Number: MAJ-1444 on . FDA classified it as Class II and gives the reason as "The MAJ-1443 and MAJ-1444 are no longer reprocessing compatible with the OER-Pro and OER-Elite automated endoscope reprocessors". As of , the recall status on record is Open, Classified.
Recall record
- Recall number
- Z-1611-2026
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- OLYMPUS Corporation of the Americas
- Product
- Olympus Endoscope Air/Water Valve - MAJ-1444. Model Number: MAJ-1444.
- Product code
- ODC Endoscope Channel Accessory
- Reason for recall
- The MAJ-1443 and MAJ-1444 are no longer reprocessing compatible with the OER-Pro and OER-Elite automated endoscope reprocessors.
- FDA root cause
- Reprocessing Controls
- Action
- Olympus notified consignees on about 02/12/2026 via URGENT: MEDICAL DEVICE CORRECTION letter. Consignees were instructed to immediately cease reprocessing the MAJ-1443 and MAJ-1444 valves in an OER-Pro or OER-Elite and ensure all personnel are completely knowledgeable and thoroughly trained on the content of this notification and the full extent of the MAJ-1443 and MAJ-1444 reprocessing steps as detailed in the provided Instructions for Use (IFU). The updated IFU can also be accessed electronically through OlympusConnect.com. Additionally, consignees were instructed to replace any existing copies of the MAJ-1443/MAJ-1444 IFU with the updated IFU provided in this letter, identify and notify all customers if affected units were further distributed and complete and confirm receipt of the notification via the Olympus web portal.
- Quantity in commerce
- 89,579 units
- Distribution
- US Nationwide distribution.
- Product codes and lots
- Model Number: MAJ-1444. UDI-DI: 04953170355929. Catalog number: N5363530. All Lot/Serial Numbers.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1610-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1611-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1612-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1613-2026 | Class II | Ongoing | Voluntary: Firm initiated |