Z-1551-2026Class II

EndoVive Safety PEG Kit; Outer Box Material Number (UPN): M00509001; Inner Box Material Number (UPN)...

Boston Scientific Corporation / initiated 2026-02-10 / Open, Classified / root cause: manufacturing process

Boston Scientific Corporation began a recall of EndoVive Safety PEG Kit; Outer Box Material Number (UPN): M00509001; Inner Box Material Number (UPN): M00509000 on . FDA classified it as Class II and gives the reason as "Recall of kits which utilized ChloraPrep Triple Swabsticks from BD. Open or compromised seals on the sterile swabstick packaging, which could result in loss of sterility or drying of the antiseptic solution. This...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1551-2026
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Boston Scientific Corporation
Product
EndoVive Safety PEG Kit; Outer Box Material Number (UPN): M00509001; Inner Box Material Number (UPN): M00509000;
Product code
PIF Gastrointestinal Tubes With Enteral Specific Connectors
Reason for recall
Recall of kits which utilized ChloraPrep Triple Swabsticks from BD. Open or compromised seals on the sterile swabstick packaging, which could result in loss of sterility or drying of the antiseptic solution. This component is placed outside of the sterile barrier of the tray manufactured by Boston Scientific, which contains the majority of the kit components. As such, inclusion of the affected ChloraPrep Triple Swabsticks does not degrade the integrity of the other components within the PEG kit.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
On February 10, 2026 Urgent Medical Device Removal - Immediate Action Required letters were sent to customers. Actions to be taken: " Do NOT use affected product (Attachment 1), remove affected product from your facility s inventory, and segregate the units in a secure place until they can be returned to Boston Scientific. " Immediately post this information in a visible location near the affected products to ensure this information is readily accessible to all handlers and users of the device. " Forward this notice to any healthcare professional from your organization for awareness and to any organization where affected devices have been transferred. " Complete and return the enclosed Reply Verification Tracking Form per the enclosed instructions. " Return affected product. This removal affects only the products and lots affected. No other material numbers or lots are impacted by this removal. If you are a facility that has sent products to another hospital or a facility within your network, ensure this notification is forwarded to them.
Quantity in commerce
806 units
Distribution
US Nationwide distribution.
Product codes and lots
Outer Box Material Number (UPN): M00509001; Inner Box Material Number (UPN): M00509000; Outer Box UDI: 08714729880394; Inner Box UDI: 08714729880387; Outer Box Lot#s: 37435463, 37522541, 37528545, 37546847, 37546850, 37546849, 37550500, 37592036, 37732292; Inner Box Lot#s: 37435462, 37514809, 37525231, 37536563, 37536564, 37536562, 37546840, 37583642, 37731897;

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1551-2026
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1551-2026Class IIOngoingVoluntary: Firm initiated