Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1695-2026Class II

Brand Name: Olympus SOLTIVE Pro SuperPulsed Laser System Product Name: SOLTIVE Pro Laser System TFL-SLS...

OLYMPUS Corporation of the Americas / initiated 2026-02-27 / Open, Classified / root cause: manufacturing process

Olympus Corporation of the Americas began a recall of Brand Name: Olympus SOLTIVE Pro SuperPulsed Laser System Product Name: SOLTIVE Pro Laser System TFL-SLS Model/Catalog Number: TFL-SLS containing TFL-CSLU... on . FDA classified it as Class II and gives the reason as "Following a complaint investigation, Olympus identified a potential defect in the 24V power supply module in some SOLTIVE laser units which may cause the system to become inoperable. Additionally, smoke or a burning...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1695-2026
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
OLYMPUS Corporation of the Americas
Product
Brand Name: Olympus SOLTIVE Pro SuperPulsed Laser System Product Name: SOLTIVE Pro Laser System TFL-SLS Model/Catalog Number: TFL-SLS containing TFL-CSLU Product Description: An electricity powered device assembly which emits a high-power laser beam to vaporize/ablate soft tissue with moderate hemostasis, little charring, and a thin zone of necrosis. Component: No
Product code
GEX Powered Laser Surgical Instrument
Reason for recall
Following a complaint investigation, Olympus identified a potential defect in the 24V power supply module in some SOLTIVE laser units which may cause the system to become inoperable. Additionally, smoke or a burning smell may occur. By design, the issue causing the smoke or burning smell would be contained within the internal laser console enclosure and would be self-extinguishing.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
On February 27, 2026 URGENT: MEDICAL DEVICE CORRECTION letters were sent to customers. Actions to be taken: 1. Examine your inventory and identify any of the affected devices listed in the attachment. 2. Olympus requests that you acknowledge receipt of this letter through our recall web portal. 3. Olympus representatives will begin contacting customers to coordinate the service of affected unit(s). In the meantime, you may continue using your SOLTIVE Laser until the service is completed. Olympus will replace the 24V power supply module at no charge to you. Note: Any issues beyond the replacement of the 24V module in this corrective action will follow the standard SOLTIVE Laser repair process and your service contract, if applicable. 4. If you have further distributed the affected products, please forward this notice to other users who may be impacted.
Quantity in commerce
32 units
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, Chile, Europe, Singapore, Australia, and Hong Kong.
Product codes and lots
Laser System: TFL-SLS; UDI: 00821925044135; Component Part: TFL-CSLU; UDI: 00821925044593; Serial #: MDUF190114, MDUF220073, MDUF220331, MDUF220353, MDUF220354, MDUF220356, MDUF220411, MDUF220412, MDUF220413, MDUF220414, MDUF220427, MDUF220451, MDUF220453, MDUF220454, MDUF220455, MDUF220456, MDUF230004, MDUF230005, MDUF230006, MDUF230007, MDUF230076, MDUF230140, MDUF230141, MDUF230163, MDUF230164, MDUF230166, MDUF230167, MDUF230234, MDUF230283, MDUF230284, MDUF230331, MDUF230332, MDUF230361, MDUF230362, MDUF230363, MDUF230364, MDUF230365, MDUF230366, MDUF230411, MDUF230412, MDUF230413;

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1695-2026
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1694-2026Class IIOngoingVoluntary: Firm initiated
Z-1695-2026Class IIOngoingVoluntary: Firm initiated