Z-1563-2026Class II

Medline Medcrest Surgical Gowns: MDTGXC4J5XL DBQ-GOWN,SURG,XALT,L4, CRTCL CVR, GRN, 5 MDTGXC4JL...

MEDLINE INDUSTRIES, LP - Northfield / initiated 2026-01-28 / Open, Classified / root cause: labeling and UDI

Medline Industries, LP began a recall of Medline Medcrest Surgical Gowns: MDTGXC4J5XL DBQ-GOWN,SURG,XALT,L4, CRTCL CVR, GRN, 5 MDTGXC4JL DBD-GOWN,SURG,XALT,L4,CRTCL CVR,GREEN,L MDTGXC4JXL... on . FDA classified it as Class II and gives the reason as "Fabric may experience premature delamination when using included laundering instructions. Identifying delaminated gown or drape in a surgical setting may result in brief delay in patient care. If delaminated product is...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-1563-2026
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
MEDLINE INDUSTRIES, LP - Northfield
Product
Medline Medcrest Surgical Gowns: MDTGXC4J5XL DBQ-GOWN,SURG,XALT,L4, CRTCL CVR, GRN, 5 MDTGXC4JL DBD-GOWN,SURG,XALT,L4,CRTCL CVR,GREEN,L MDTGXC4JXL DBD-GOWN,SURG,XALT,L4,CRTCL CVR,GREEN,XL MDTGXC4JXXL DBD-GOWN,SURG,XALT,L4,CRTCL CVR,GREEN,XX MDTGXP4JL DBD-GOWN,SURG,XALT,L4,PANEL CVR,GREEN,L MDTGXP4JXL DBD-GOWN,SURG,XALT,L4,PANEL CVR,GREEN,XL MDTGXP4JXXL DBD-GOWN,SURG,XALT,L4,PANEL CVR,GREEN,XX MDTL1002862EF DBD-GOWN,SURG,XALT,L4,PANEL CVR,GREEN,XL MDTZ1000893AJ DBD-GOWN,SRG,PNL CVR, GREEN,SNPS,XL
Product code
FYA Gown, Surgical
Reason for recall
Fabric may experience premature delamination when using included laundering instructions. Identifying delaminated gown or drape in a surgical setting may result in brief delay in patient care. If delaminated product is used and surgical strikethrough occurs, there may be risk of infection.
Root cause tag
labeling and UDI
FDA root cause
Labeling Change Control
Action
On January 28, 2026, the firm notified customers of the issue through Urgent Medical Device Recall letters. Users were provided with updated laundering instructions and general inspection guidelines.
Quantity in commerce
104,004
Distribution
Domestic distribution nationwide US. International distribution to Panama, United Arab Emirates, Virgin Islands, Saudi Arabia, India, Chile, and Canada.
Product codes and lots
UDI-DI (each/case) 10197344014002 20197344014009 10888277189188 40888277189189 10884389189191 40884389189192 10884389193297 40884389193298 10888277189218 40888277189219 10884389189221 40884389189222 10193489040111 20193489040118 10884389189221 40884389189222 10193489021820 20193489021827 All lots affected

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1563-2026
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1563-2026Class IIOngoingVoluntary: Firm initiated
Z-1564-2026Class IIOngoingVoluntary: Firm initiated