Medline Surgical Drapes: MDTBTCS6070GN DBD-DRAPE,COVER,TABLE,RESIS/XALT GRN,60X MDTBTCS6090GN...
Medline Industries, LP began a recall of Medline Surgical Drapes: MDTBTCS6070GN DBD-DRAPE,COVER,TABLE,RESIS/XALT GRN,60X MDTBTCS6090GN DBD-DRAPE,COVER,TABLE,RESIS/XALT GRN,60X MDTDXUNDBGN... on . FDA classified it as Class II and gives the reason as "Fabric may experience premature delamination when using included laundering instructions. Identifying delaminated gown or drape in a surgical setting may result in brief delay in patient care. If delaminated product is...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-1564-2026
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- MEDLINE INDUSTRIES, LP - Northfield
- Product
- Medline Surgical Drapes: MDTBTCS6070GN DBD-DRAPE,COVER,TABLE,RESIS/XALT GRN,60X MDTBTCS6090GN DBD-DRAPE,COVER,TABLE,RESIS/XALT GRN,60X MDTDXUNDBGN DBD-DRAPE,UNDERBUTTOCK,28 X 50.5,XALT GR MDTMAYOXGN DBD-DRAPE,COVER,MAYO STAND,RESISTAT/XALT MDTZ1003656GN DBD-POCKET SHEET,RESISTAT/XALT
- Product code
- PUI Drape, Surgical, Exempt
- Reason for recall
- Fabric may experience premature delamination when using included laundering instructions. Identifying delaminated gown or drape in a surgical setting may result in brief delay in patient care. If delaminated product is used and surgical strikethrough occurs, there may be risk of infection.
- Root cause tag
- labeling and UDI
- FDA root cause
- Labeling Change Control
- Action
- On January 28, 2026, the firm notified customers of the issue through Urgent Medical Device Recall letters. Users were provided with updated laundering instructions and general inspection guidelines.
- Quantity in commerce
- 23,568
- Distribution
- Domestic distribution nationwide US. International distribution to Panama, United Arab Emirates, Virgin Islands, Saudi Arabia, India, Chile, and Canada.
- Product codes and lots
- UDI-DI (each/case) 10193489036961 20193489036968 10193489036978 20193489036975 10193489040098 20193489040095 10193489036954 20193489036951 10193489036985 20193489036982 All lots affected
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1563-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1564-2026 | Class II | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.