Brand Name: Welch Allyn, Inc. Product Name: Welch Allyn Spot Vision Screener VS100 Model/Catalog Number...
Baxter Healthcare Corporation began a recall of Brand Name: Welch Allyn, Inc. Product Name: Welch Allyn Spot Vision Screener VS100 Model/Catalog Number: VS100 Software Version: Software version 3.2.0.1... on . FDA classified it as Class II and gives the reason as "Display screen may unintentionally flicker due to a software issue. Exposure to flickering lights may induce photosensitive seizure activity in susceptible individuals". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1573-2026
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Baxter Healthcare Corporation
- Product
- Brand Name: Welch Allyn, Inc. Product Name: Welch Allyn Spot Vision Screener VS100 Model/Catalog Number: VS100 Software Version: Software version 3.2.0.1 Product Description: Welch Allyn Spot Vision Screener VS100, FDA product code name: Refractometer, ophthalmic One unit per package
- Product code
- HKO Refractometer, Ophthalmic
- Reason for recall
- Display screen may unintentionally flicker due to a software issue. Exposure to flickering lights may induce photosensitive seizure activity in susceptible individuals.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- On February 18, 2026, the firm notified consignees via Urgent Medical Device Recall letters. Users were instructed to immediately locate, isolate, and stop using the listed affected products. Customers should contact Baxter Technical Support to arrange for the return and correction of their impacted device. Baxter may provide a loaner device upon request while your device is being upgraded and serviced. Distributors should return unopened products in inventory for credit. If affected product was further distributed, the distributor should conduct a consumer-level recall.
- Quantity in commerce
- 59
- Distribution
- Worldwide - US Nationwide distribution in the states of AZ, CO, IL, IN, KS, MA, NE, NJ, NM, NY, PA, SC, TX, VA and the countries of Colombia and India.
- Product codes and lots
- Lot Code: " Product Code: HKO " Version or Model: 901029 " Catalog Number: VS100 " Company Name: WELCH ALLYN, INC. " Primary DI Number: 00732094214154 " Packaging DI Number: 00732094212228 " Device Description: VS100 UDI BASE UNIT " Software version: 3.2.0.1 " Serial Numbers: Refer to attached Affected Product Table
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1573-2026 | Class II | Ongoing | Voluntary: Firm initiated |