Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1573-2026Class II

Brand Name: Welch Allyn, Inc. Product Name: Welch Allyn Spot Vision Screener VS100 Model/Catalog Number...

Baxter Healthcare Corporation / initiated 2026-02-18 / Open, Classified / root cause: design

Baxter Healthcare Corporation began a recall of Brand Name: Welch Allyn, Inc. Product Name: Welch Allyn Spot Vision Screener VS100 Model/Catalog Number: VS100 Software Version: Software version 3.2.0.1... on . FDA classified it as Class II and gives the reason as "Display screen may unintentionally flicker due to a software issue. Exposure to flickering lights may induce photosensitive seizure activity in susceptible individuals". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.

Recall record

Recall number
Z-1573-2026
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Baxter Healthcare Corporation
Product
Brand Name: Welch Allyn, Inc. Product Name: Welch Allyn Spot Vision Screener VS100 Model/Catalog Number: VS100 Software Version: Software version 3.2.0.1 Product Description: Welch Allyn Spot Vision Screener VS100, FDA product code name: Refractometer, ophthalmic One unit per package
Product code
HKO Refractometer, Ophthalmic
Reason for recall
Display screen may unintentionally flicker due to a software issue. Exposure to flickering lights may induce photosensitive seizure activity in susceptible individuals.
Root cause tag
design
FDA root cause
Device Design
Action
On February 18, 2026, the firm notified consignees via Urgent Medical Device Recall letters. Users were instructed to immediately locate, isolate, and stop using the listed affected products. Customers should contact Baxter Technical Support to arrange for the return and correction of their impacted device. Baxter may provide a loaner device upon request while your device is being upgraded and serviced. Distributors should return unopened products in inventory for credit. If affected product was further distributed, the distributor should conduct a consumer-level recall.
Quantity in commerce
59
Distribution
Worldwide - US Nationwide distribution in the states of AZ, CO, IL, IN, KS, MA, NE, NJ, NM, NY, PA, SC, TX, VA and the countries of Colombia and India.
Product codes and lots
Lot Code: " Product Code: HKO " Version or Model: 901029 " Catalog Number: VS100 " Company Name: WELCH ALLYN, INC. " Primary DI Number: 00732094214154 " Packaging DI Number: 00732094212228 " Device Description: VS100 UDI BASE UNIT " Software version: 3.2.0.1 " Serial Numbers: Refer to attached Affected Product Table

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1573-2026
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1573-2026Class IIOngoingVoluntary: Firm initiated