Z-1586-2026Class II
Brand Name: Aptima HPV Assay Product Name: Aptima HPV Assay Model/Catalog Number: 303585 Product Description...
Hologic, Inc. / initiated 2026-02-18 / Open, Classified
Hologic, Inc. began a recall of Brand Name: Aptima HPV Assay Product Name: Aptima HPV Assay Model/Catalog Number: 303585 Product Description: Aptima HPV assay Component: No on . FDA classified it as Class II and gives the reason as "Due to product exhibiting potential to generate either invalid or false negative results". As of , the recall status on record is Open, Classified.
Recall record
- Recall number
- Z-1586-2026
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Hologic, Inc.
- Product
- Brand Name: Aptima HPV Assay Product Name: Aptima HPV Assay Model/Catalog Number: 303585 Product Description: Aptima HPV assay Component: No
- Product code
- OYB Kit, Rna Detection, Human Papillomavirus
- Reason for recall
- Due to product exhibiting potential to generate either invalid or false negative results.
- FDA root cause
- Under Investigation by firm
- Action
- On 02/18/2026, the firm sent via FedEx an "URGENT: MEDICAL DEVICE RECALL NOTIFICATION" letter to customers informing them that Hologic has identified that the Aptima HPV Assay kits from ML 929785 exhibit the potential to generate either invalid results for samples and calibrators or false negative results for low analyte samples when reagents are not used on the day of reconstitution. Customers are instructed: 1. Discontinue use of the Aptima HPV Assay kits from ML 929785. 2. Segregate and count the number of affected product on-hand and dispose according t the local guidelines and regulations. 3.Fill out the Customer Response Form using the link https://iqvia-response.my.site.com/mt/fca?cid=hpv-929785 4. Hologic will process and ship the replacement kits For Questions or concerns, contact Hologic Technical Support at +1 888-484-4747 or +1 858-410-8511 during hours 5 am to 5 pm PT or by email at [email protected]
- Quantity in commerce
- 306
- Distribution
- U.S. Nationwide distribution in the states of CA, FL. IA, IL, MO, NC, NM, PR, and TN.
- Product codes and lots
- Lot Code: The Aptima HPV Assay Kit Part Number 303585, Master Lot 929785 contains one Refrigerated Box (PN 302883) and one Room Temperature Box (PN 302885). The following are the UDI number for each box: UDI Number for PN 302883: 15420045500051 UDI Number for PN 302885: 15420045500068
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1586-2026 | Class II | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.