Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1586-2026Class II

Brand Name: Aptima HPV Assay Product Name: Aptima HPV Assay Model/Catalog Number: 303585 Product Description...

Hologic, Inc. / initiated 2026-02-18 / Open, Classified

Hologic, Inc. began a recall of Brand Name: Aptima HPV Assay Product Name: Aptima HPV Assay Model/Catalog Number: 303585 Product Description: Aptima HPV assay Component: No on . FDA classified it as Class II and gives the reason as "Due to product exhibiting potential to generate either invalid or false negative results". As of , the recall status on record is Open, Classified.

Recall record

Recall number
Z-1586-2026
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Hologic, Inc.
Product
Brand Name: Aptima HPV Assay Product Name: Aptima HPV Assay Model/Catalog Number: 303585 Product Description: Aptima HPV assay Component: No
Product code
OYB Kit, Rna Detection, Human Papillomavirus
Reason for recall
Due to product exhibiting potential to generate either invalid or false negative results.
FDA root cause
Under Investigation by firm
Action
On 02/18/2026, the firm sent via FedEx an "URGENT: MEDICAL DEVICE RECALL NOTIFICATION" letter to customers informing them that Hologic has identified that the Aptima HPV Assay kits from ML 929785 exhibit the potential to generate either invalid results for samples and calibrators or false negative results for low analyte samples when reagents are not used on the day of reconstitution. Customers are instructed: 1. Discontinue use of the Aptima HPV Assay kits from ML 929785. 2. Segregate and count the number of affected product on-hand and dispose according t the local guidelines and regulations. 3.Fill out the Customer Response Form using the link https://iqvia-response.my.site.com/mt/fca?cid=hpv-929785 4. Hologic will process and ship the replacement kits For Questions or concerns, contact Hologic Technical Support at +1 888-484-4747 or +1 858-410-8511 during hours 5 am to 5 pm PT or by email at [email protected]
Quantity in commerce
306
Distribution
U.S. Nationwide distribution in the states of CA, FL. IA, IL, MO, NC, NM, PR, and TN.
Product codes and lots
Lot Code: The Aptima HPV Assay Kit Part Number 303585, Master Lot 929785 contains one Refrigerated Box (PN 302883) and one Room Temperature Box (PN 302885). The following are the UDI number for each box: UDI Number for PN 302883: 15420045500051 UDI Number for PN 302885: 15420045500068

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1586-2026
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1586-2026Class IIOngoingVoluntary: Firm initiated